7. Maintenance Schedule and Daily Check List
All Oxford products are designed for minimum maintenance, however some safety checks and
procedures are required. A schedule of DAILY tasks are detailed below.
Daily checks and a bi-annual service, inspection and test will ensure your lift is kept in optimum
safe working condition. A list of spare parts is available upon request.
NOTE:
The LOAD TeST and CerTIfICATION should only be carried out by qualified personnel
or an authorised service dealer.
Daily Check List:
Joerns Healthcare strongly recommend the following checks are carried out on a daily basis and
before using the lift.
• ENSURE
the lift moves freely on its castors.
• ENSURE
the legs open and close correctly.
• EXAMINE
the sling hooks for excessive wear and
ENSURE
the retainers are present and
functioning as intended. If in doubt,
DO NOT
use.
•
Operate the hand control to
CONFIRM
the boom raises and lowers satisfactorily.
• CONFIRM
the lift is not giving a low battery alarm when the hand control is operated. If the
alarm sounds,
DO NOT
use and place on charge immediately.
• CHECK
the operation of the emergency stop button.
• ENSURE
the mast is fully engaged and the locking knob is fully tightened.
• EXAMINE
slings for fraying or other damage.
DO NOT
use any sling if damaged or if the
sling shows signs of wear.