14
English
Hoyer
®
HPL500
8.
Emergency lower:
In the case of a complete electrical failure the electrical actuator
is fitted with a mechanical lowering device (RED BOSS). The device must be twisted
clockwise by hand to operate. A slow descent will commence. Repeat this process until
the patient has been safely lowered.
WARNING
The emergency lowering device should only be used in the event of electrical
failure. It must not be used to lower the patient on a regular basis.
WARNING
If this feature is used, the lift MUST be subsequently checked out by a competent
technician.
WARNING
Continued use of the lift after using the emergency descent feature may
compromise the duty of care and safety of the patient.
9. Batteries:
The batteries are protected from deep discharge by a LOW VOLTAGE
ALARM. This will sound when the batteries need recharging and the hand control is being
operated. It will not sound independently of the hand control being operated. DO NOT
IGNORE THIS WARNING ALARM. Complete the lifting operation and place the battery on
charge (see charging instructions).
10. Slings:
The selected sling is attached to the spreader bar hooks. Each sling is supplied
with instructions which should be followed carefully.
The Hoyer HPL500 is suitable for patients in the SITTING, SITTING/RECUMBENT and
RECUMBENT positions. The core slings suitable for these lifts are listed as follows:
• Hoyer Full Back
• Hoyer Access
• Hoyer Quickfit
• Hoyer Long Seat
• Hoyer Quickfit Deluxe
When selecting a sling from the Hoyer range, be sure to assess the suitability of the type of
sling for the patient to be lifted by conducting a thorough risk assessment prior to lifting.
NOTE:
For detailed fitting instructions, please refer to the user guide supplied with each sling.
Fig. 17