background image

User/Service Manual 

for 

Joerns

®

Easy Care

®

2003DC

Model B684DC

To avoid injury, read user’s manual before using bed.

Содержание Easy Care 2003DC

Страница 1: ...User Service Manual for Joerns Easy Care 2003DC Model B684DC To avoid injury read user s manual before using bed...

Страница 2: ...esident mattress bedding and any other equipment or persons likely to be on the bed Do not exceed the 500 pound capacity or use water mattresses that allow water to flow to the end of the bed when the...

Страница 3: ...will allow a resident to get his or her head or neck lodged within these openings Failure to do so could result in injury or death WARNING POSSIBLE INJURY OR DEATH Do not use any Joerns assist device...

Страница 4: ...regulatory guidelines can be complex That s why Joerns offers a wide array of compliance options We will assist customers in selecting compliant accessories recommended for their model beds Creating a...

Страница 5: ...ds distributed evenly on the sleep surface and a normal working load of 300 pounds This is the total weight of the resident mattress bedding accessories and other medical equipment in use Do not use w...

Страница 6: ...sticker side up four holes should be visible The foot panel brackets are the shorter of the two sizes fig 4 Install the foot panel brackets with the four long screws provided The open side of the brac...

Страница 7: ...inate surfaces should be maintained by cleaning with a disinfectant spray Head Operation 1 Push button to raise Head 2 Push button to lower Head Knee Operation 3 Push button to raise Knee 4 Push butto...

Страница 8: ...l cords are plugged in including pendant staff control cords and AC Power 2 There is a protection feature built into the motors and control box to prevent overheating If the bed was operating continuo...

Страница 9: ...e gap between the mattress and the head foot board is small enough to prevent a resident from getting his or her head or neck caught in these locations Failure to do so could result in injury or death...

Страница 10: ...ay cause it to disconnect from the bed resulting in a loss of control of the bed 4 Extra care should be taken when moving the bed on any slope Moving the bed on a slope requires two people one at each...

Страница 11: ...harger Green light indicates main power is ON Yellow light indicates battery is charging After approximately 4 hours of charging the yellow light goes off and the battery will be fully charged To Powe...

Страница 12: ...dical will void this warranty Parts Joerns beds contain various parts that wear from normal use These parts such as DC batteries are not covered under the 2 year warranty but are covered for 90 days a...