PAGE 44 of 53
Chapter 9
TROUBLESHOOTING
PROBLEM
POSSIBLE CAUSE
SOLUTION
Chain is too
loose/tight.
Incorrect number or
combination of links.
Refer to the
CIRC-IT
setup chart
(see CIRC-IT
Setup Chart on page 53)
to assemble the
correct configuration for the size of pipe/tube
being scanned.
Incorrect latch setting. Ensure the latch is set using the tightest notch
possible
(see Latch Connection on page 40).
Incorrect latch used.
When scanning pipe/tube with an OD less than
50.8 mm
(2 in)
, the short latch must be used.
When scanning pipe/tube with an OD greater
than 50.8 mm
(2 in)
, the long latch must be used
(see CIRC-IT Setup Chart on page 53).
Insufficient probe
contact.
Scanner not set
properly.
Reconfigure the scanner as per instructions
(see
Setup of CIRC-IT on a Pipe or Tube on page 14).
Pivot adjustment
screws were not
adequately tightened.
Reconfigure the scanner as per instructions
(see
Setup of CIRC-IT on a Pipe or Tube on page 14)
and ensure pivot adjustment screws are tight.
Setup tab not raised.
Ensure the setup tab has been raised and the
screws tightened.
(see step 13 on page 18).
Setup tab are located on each side of a dual
probe scanner.
Cables and/or hoses
are routed incorrectly.
Ensure the cables and hoses are not impeding
proper scanner operation.
Clearance issues
when scanning
small diameters.
Remove encoder
cable guard.
(see Setup on diameters smaller than 31.75 mm (1.25
in) on page 18)
9.1. Technical Support
For technical support contact Jireh Industries
(see Jireh Industries Ltd. on page 1).