![Jinko JK2520B Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/jinko/jk2520b/jk2520b_user-manual_2018511043.webp)
43
JK2520C AC resistance indicator:
RANGE
Max
Reading
Resoluti
on
ULTRA
FAST
MED
SLOW
Test
Current
1
30mΩ 33.000mΩ 1μΩ
1.5% ± 20
1.25%
± 10
1% ± 10 1% ± 5
150mA
2
300m
Ω
330.00mΩ 10μΩ
1.5% ± 20
1.25%
± 10
1% ± 10 1% ± 5
15mA
3
3Ω
3.3000Ω
100μΩ
1.5% ± 20
1.25%
± 10
1% ± 10 1% ± 5
1.5mA
4
30Ω
33.000Ω
1mΩ
1.5% ± 20
1.25%
± 10
1% ± 10 1% ± 5
150μA
5
300Ω 330.00Ω
10mΩ
1.5% ± 20
1.25%
± 10
1% ± 10 1% ± 5
15μA
6
3kΩ
3.3000kΩ
100mΩ
1.5% ± 20
1.25%
± 10
1% ± 10 1% ± 5
15μA
7
30kΩ
33.000kΩ
1Ω
1.8% ± 50
1.25%
± 20
1% ± 20 1% ±
10
1.5μA
JK2520C DC voltage indicator:
RANGE
Max
Reading
Resolution ULTRA
FAST
MED
SLOW
0
6V
12.000V
100μV
0.5%±100 0.3%±50
0.2%±10
0.2%±5
1
60V
120.000V
1mV
0.5%±100 0.3%±50
0.2%±10
0.2%±5
JK2520B AC resistance indicator:
RANGE
Max Reading Resolution
FAST
SLOW
Test Current
1
10mΩ
30.00mΩ
10μΩ
1.5% ± 10
1% ± 5
150mA
2
100mΩ
300.00mΩ
100μΩ
1.5% ± 10
1% ± 5
15mA
3
1Ω
3.00Ω
1mΩ
1.5% ± 10
1% ± 5
1.5mA
4
10Ω
30.00Ω
10mΩ
1.5% ± 10
1% ± 5
150μA
JK2520B DC voltage indicator
:
RANGE
Max Reading
Resolution
FAST
SLOW
0
3V
6.00V
1mV
0.5%±50
0.3%±5
1
30V
60.00V
10mV
0.5%±50
0.3%±5
10.2 General specification
Screen: TFT-LCD true color display, screen size 4.3 inches.
Test speed:0.--- manual range:
Slow: 1 times / sec.
Medium speed: 10 times / sec
Fast: 20 times / sec.