![Jinko JK2520B Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/jinko/jk2520b/jk2520b_user-manual_2018511039.webp)
39
Command grammar
:
COMParator:TOLerance:RNOM <float>
For example: Send out > COMP:TOL:NOM 1m
//The nominal value is set to 1mΩ
Send out > COMP:TOL:NOM 1E-3
//The nominal value is set to 1mΩ
Send out > COMP:TOL:NOM 1000
//The nominal value is set to 1mΩ
Query syntax
:
COMP:TOL:RNOM?
Query response
:
<scifloat>
For example: Send out > COMP:TOL:NOM? <NL>
Return > 1.0000E-03<NL>
//The nominal value is set to 1mΩ
9.6.5 COMParator:TOLerance:VNOMinal
COMP:TOL:VNOM Used to set the voltage nominal value
Command grammar
:
COMParator:TOLerance:VNOM <float>
For example: Send out > COMP:TOL:VNOM 1.23 //The nominal value is set to 1.23V
Send out > COMP:TOL:VNOM 50
//The nominal value is set to 50V
Query syntax
:
COMP:TOL:VNOM?
Query response
:
<scifloat>
For example:
Send out > COMP:TOL:VNOM? <NL>
Return > 1.0000E+00<NL>
// nominal value is 1V
9.6.6
COMParator:TOLerance:RLMT
COMP:TOLerance:RLMT Used to set the nominal value
Command grammar
:
COMParator:TOLerance:RLMT <float lower limit>,<float Upper
limit>
For example: Send out > COMP:TOL:RLMT -10,+10 //If the percentage is selected, the
lower limit is -10% and the upper limit is 10%.
Query syntax
:
COMP:TOL:RLMT?
Query response
:
<scifloat>,<scifloat>
For example: Send out > COMP:Beep? 1<NL>
Return > -10.000E+00,+10.000E+00<NL> //-10,+10
9.6.7
COMParator:TOLerance:VLMT
COMP:TOLerance:VLMT Used to set the nominal value
Command grammar
:
COMParator:TOLerance:VLMT <float lower limit>,<float Upper limit>
For example: Send out > COMP:TOL:VLMT -10,+10 //If the percentage is selected, the
lower limit is -10% and the upper limit is 10%.
Query syntax
:
COMP:TOL:VLMT?
Query response
:
<scifloat>,<scifloat>
For example: Send out > COMP:TOL:VLMT?<NL>
Return > -10.000E+00,+10.000E+00<NL> //-10,+10
9.7 TRIGger subsystem
Fig 9
‐
5
TRIGger subsystem tree