![Jinko JK2520B Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/jinko/jk2520b/jk2520b_user-manual_2018511015.webp)
15
Trigger – trigger mode
Range - Test Range
Speed - test speed
Alarm - comparator alarm alarm
Figure 4 -1 < measurement display> page
Slow, medium speed and fast mode
High speed mode
4.1.1
[Trigger] mode
SCPI Command: TRIGger:SOURce {INT,MAN,EXT,BUS}
SCPI Query Command: TRIGger:SOURce?
The instrument has 4 triggering modes:
Internal triggering, manual triggering, external triggering and remote triggering.
Trigger
mode
Describe
Internal
Also known as continuous testing, triggering signals from inside the
instrument Testing continuously according to inherent cycle.
manual
Every time the [Trig] key is pressed, the instrument performs a
measuring cycle.Other times the instrument is in wait state.
external
An ascending edge pulse is received from the Handler interface of
the rear panel, and the instrument performs a measuring cycle. The
other time is waiting. Please refer to the Handler interface.
remote
After receiving the RS232 trigger command, the instrument executes
one measurement cycle and the other time is in a waiting state.
* set trigger mode steps:
The first step is to press the Meas key into the main page.
The second step is to use the cursor key to select the trigger field.
The third step is to use function keys to select trigger mode.
Function key
Function