13
Chapter 5
About accuracy
5.
1
For the whole measuring range, the accuracy complies with the curve below.
20.0
Accuracy(
%)
10kHz
1kHz
10.0
5.0
2.0
1.0
100Hz
0.5
0.2
0.1
0 .01mH
.1mH
1mH
10mH 0.1H
1H
10H
100H
1000H
H
5.2
. Accuracy for Quality Factor in the whole range: ±[0.030+0.2(1+Q)%]
5.3.
Accuracy for Dissipation in the whole range: ±[0.0030+0.5D(1+D)%]
Chapter 6 Maintenance and
Notice
6
.
1.
The meter should connect with earth reliably.
6.2.
If this meter displays the chaotic or disordered data, users can recover it to factory
Содержание 2811C