TLS5 User Manual
TLS5-UM-00001 Rev 002
36
Standard Event Status Register Query
Syntax
*ESR?
Function
Returns the contents of the standard event status register as an
integer that, when converted to a binary number, represents the bit
values of the register. The standard event status register is cleared
after
*ESR?
command.
Example
*ESR?
returns 195 if the standard event status register is set to
11000011.
Identification Query
Syntax
*IDN?
Function
The
*IDN
query returns a string value which identifies the
manufacturer, instrument type and firmware version.
Example
*IDN?
Returns "JGR Optics Inc., BR5, XXXXXXX, Y.YY"
Where:
<XXXXXXX> = device serial number
<Y.YY> = firmware revision number
Operation Complete Command
Syntax
*OPC
Function
Causes the source to set the OPC bit in the standard event status
register when all pending operations have been completed.
Example
*OPC