Technical data
61
10.4 Setting
ranges
and
accuracy
of
the
alarm
parameters
10.5 Pneumatic
circuit
diagram
F(C): Flow meter with regulator characteristics if target volume or target min-
ute volume are set.
PC: Pressure measurement with regulator characteristics.
Alarm
Setting range
Accuracy
High respiratory rate
Range: Off, 10-60 bpm
Step:1 bpm
± 1 bpm
Low respiratory rate
Rate: Off, 4-20 bpm
Step: 1 bpm
± 1 bpm
Low minute volume
Range: Off, 2.000-10.000 l
Step: 0.100 l
± 0.2 Pa
Leakage
Range: On, off
Not necessary
Non vented mask
Range: On, off
Not necessary
Target volume not reached
Range: On, off
± 0.2 Pa
Target minute volume not
reached
Range: On, off
± 0.2 Pa