Technical data
59
10.3 Setting
ranges
and
accuracy
of
the
therapy
settings
Parameter
Setting range
Accuracy
CPAP
Range: 3-20 hPa/cm H
²
O
Step: 0.5 hPa/cm H
²
O
± 0.4 hPa
IPAP
Range: 5-30 hPa/cm H
²
O
Step: 0.5 hPa/cm H
²
O
Dependency:
≥
EPAP + 2 hPa/cm H
²
O
± 0.4 hPa
EPAP
Range: 3-25 hPa/cm H
²
O
Step: 0.5 hPa/cm H
²
O
Dependency:
≤
IPAP - 2 hPa/cm H
²
O
± 0.4 hPa
Target volume
Range: Off, 0.100-2.000 l
Step: 0.100 l
± 0.2 Pa
Target minute vol-
ume
Range: Off, 2.000-10.000 l
Step: 0.100 l
± 0.2 Pa
Pressure support
Range: 0-25 hPa cm H
²
O
Step: 0.5 hPa/cm H
²
O
Dependency:
≤
30 - IPAP
± 0.4 hPa
Maximum pres-
sure increase per
breath
Range: 0.1-2 hPa/cm H
²
O
Step: 0.1 hPa/cm H
²
O
± 0.4 hPa
Inspiration slope
Range: Step 1-6
Step: 1
± 0.4 hPa
Expiration slope
Range: Step 1-6
Step: 1
± 0.4 hPa
Inspiration trigger
Range: Step 1-5
Step: 1
± 0.2 Pa
Expiration trigger
Range: Auto, 30-70%
Step: 10%
± 0.2 Pa
Trigger lock
Range: 0.3-8.0 s
Step: 0.1 s
Dependency:
≤
(60:respiratory rate)-
inspiration time
± 1 ms