5. Calibration coefficients:
Coefficients
C0=F
C1=F
C2=F
d0=F
d1=F
d2=F
d3=F
d4=F
Cp=F
e0=F
FilmNo=S
docaldate=S
A=F
B=F
C=F
D=F
E=F
F=F
G=F
H=F
tcaldate=S
Example of command for calibration coefficients:
dc,OK,46,<CR><LF>
C0=3.24923E-03,E9,<CR><LF>
C1=1.40598E-04,E5,<CR><LF>
…
docaldate=2013/05/21,09,<CR><LF>
A=2.345,59,<CR><LF>
B=13.66666,B4,<CR><LF>
…
tcaldate=2013/05/21,68,<CR><LF>
6. Data format
HEX data format for commands
do
,
tdo
,
sdo
, and
stdo
;
The HEX format is extracted as follows.
- Temperature
HEX=TTTT
TTTT is a value for temperature and four-letter HEX value.
Temperature = (decimal expression for TTTT) / 1000 - 5
If temperature < -5.000, then set 0000 to TTTT.
If temperature > 40.000, then set FFFF to TTTT.
- DO
HEX=DDDD
DDDD is a value for DO and four-letter HEX value.
DO = (decimal expression for DDDD) / 100.
If DO > 425.00, then set FFFF to DDDD.
HEX data format for commands
tdon
,
tdona
,
stdon
, and
stdona
;
The HEX format is extracted as follows.
- Temperature
HEX is between 0000 and FFFF.
Fixed length of 4
- DO
HEX is between 0000 and FFFF.
Fixed length of 4
- LED accumulated time
HEX is between 00000000 and FFFFFFFF.
Fixed length of 8
The LED accumulated time is expressed by a factor of 10 ms, e.g., the output value “5” means 50 ms.
In order to calculate the LED accumulated time in seconds,
LED accumulated time (s) = (decimal expression for LED accumulated time HEX) / 100
“F” data format
“F” is a real number.
“F” is variable-length and 16 at maximum.
Temperature C coefficient. “F” is a real number.
Temperature B coefficient. “F” is a real number.
Temperature A coefficient. “F” is a real number.
Date of calibration for temperature sensor. The format of “S” is “YYYY/MM/DD”.
Temperature H coefficient. “F” is a real number.
Temperature G coefficient. “F” is a real number.
Temperature D coefficient. “F” is a real number.
d3 coefficient (Uchida
et al.
, 2013). “F” is a real number.
d4 coefficient (Uchida
et al.
, 2013). “F” is a real number.
Cp coefficient (Uchida
et al.
, 2013). “F” is a real number.
e0 coefficient (Uchida
et al.
, 2013). “F” is a real number.
Film number “S” is an ASCII string [0-9a-zA-Z]. The length of the string is 8.
Date of calibration for DO sensor. The format of “S” is “YYYY/MM/DD”.
Description
C0 coefficient (Uchida
et al.
, 2013). “F” is a real number.
C1 coefficient (Uchida
et al.
, 2013). “F” is a real number.
C2 coefficient (Uchida
et al.
, 2013). “F” is a real number.
d0 coefficient (Uchida
et al.
, 2013). “F” is a real number.
d1 coefficient (Uchida
et al.
, 2013). “F” is a real number.
d2 coefficient (Uchida
et al.
, 2013). “F” is a real number.
Temperature F coefficient. “F” is a real number.
Temperature E coefficient. “F” is a real number.
[Appendix 1] 3 /4
Содержание RINKO FT
Страница 2: ......
Страница 36: ...34 Dimensions are in mm Figure 9 3 Cable type ARO FT Figure 9 4 MCBH type AROD FT...
Страница 39: ......