4. Commands:

HC to RINKO FT RINKO FT to HC Description

Request

Response

Outline

Detail

do

do,HEX

Send DO,

enter sleep mode.

Turn ON the analog power (if it is in sleep mode), send the DO physical data [μ

mol/L], and enter sleep mode. If the RINKO FT analog power is OFF when receiving

the command, it takes 5 seconds for the DO value to stabilize. For that reason, DO

NOT USE this command during the sleep mode because the RINKO FT again enters

sleep mode immediately after sending unstable/unreliable data.

tdo

tdo,HEX,HEX

Send T and DO,

enter sleep mode.

Turn ON the analog power (if it is in sleep mode), send the temperature and DO

physical data [

°

C and μmol/L], and enter sleep mode. If the RINKO FT analog power

is OFF when receiving the command, it takes 5 seconds for the DO value to stabilize.

For that reason, DO NOT USE this command during the sleep mode because the

RINKO FT again enters sleep mode immediately after sending unstable/unreliable

data. The left and right HEXs represent temperature and DO, respectively.

sdo

sdo,HEX

Send DO.

Turn ON the analog power (if it is in sleep mode) and send the DO physical data [μ

mol/L]. If the RINKO FT analog power is OFF when receiving the command, it takes

5 seconds for the DO value to stabilize.

stdo

stdo,HEX,HEX

Send T and DO.

Turn ON the analog power (if it is in sleep mode) and send the temperature and DO

physical data [°C and μmol/L]. If the RINKO FT analog power is OFF when

receiving the command, it takes 5 seconds for the DO value to stabilize. The left and

right HEXs represent temperature and DO, respectively.

tdon

tdon,HEX,HEX,H

EX

Send AD values of T and DO

and LED time,

enter sleep mode.

Turn ON the analog power (if it is in sleep mode), and send the temperature and DO

AD values and LED accumulated time, and enter sleep mode. If the RINKO FT

analog power is OFF when receiving the command, it takes 5 seconds for the DO

value to stabilize. For that reason, DO NOT USE this command during the sleep

mode because the RINKO FT again enters sleep mode immediately after sending

unstable/unreliable data. The first HEX represents the temperature, the second the

DO, and the third the LED accumulated time.

tdona

tdona,HEX,HEX,

HEX,HEX,HEX,

HEX,HEX

Send all parameters as AD

values and LED time,

enter sleep mode.

Turn ON the analog power (if it is in sleep mode), send temperature and DO AD

values, phases of the blue and red lights (AD values), amplitudes of the blue and red

lights (AD values), and LED accumulated time, and then enter sleep mode. If the

RINKO FT analog power is OFF when receiving the command, it takes 5 seconds for

the DO value to stabilize. For that reason, DO NOT USE this command during the

sleep mode because the RINKO FT again enters sleep mode immediately after

sending unstable/unreliable data. The first HEX represents the temperature, the

second the DO, the third the blue phase, the fourth the red phase, the fifths the blue

amplitude, the sixth the red amplitude, and the seventh the LED accumulated time.

stdon

stdon,HEX,HEX,

HEX

Send AD values of T and DO

and LED time.

Turn ON the analog power (if it is in sleep mode) and send the temperature and DO

AD values and LED accumulated time. If the RINKO FT analog power is OFF when

receiving the command, it takes 5 seconds for the DO value to stabilize. The first

HEX represents the temperature, the second the DO, and the third the LED

accumulated time.

stdona

stdona,HEX,HEX,

HEX,HEX,HEX,

HEX,HEX

Send all parameters as AD

values and LED time.

Turn ON the analog power (if it is in sleep mode) and send temperature and DO AD

values, phases of the blue and red lights (AD values), amplitudes of the blue and red

lights (AD values), and LED accumulated time. If the RINKO FT analog power is

OFF when receiving the command, it takes 5 seconds for the DO value to stabilize.

The first HEX represents the temperature, the second the DO, the third the blue

phase, the fourth the red phase, the fifths the blue amplitude, the sixth the red

amplitude, and the seventh the LED accumulated time.

qs

qs,OK

Enter sleep mode.

Enter sleep mode after returning a response. When turning off the power of the

RINKO FT, do so after the RINKO FT is in sleep mode.

*serialnumber

*serialnumber=S Display serial number

The serial number of the RINKO FT. “S” is an ASCII string [0-9a-zA-Z]. The length

of the string is 10 at maximum.

dc

(1)dc,OK

(2)

Coefficients

Display calibration

coefficients.

(1) First, it gives a response to the command (

Response

=dc,OK).

(2) Next, returns the calibration 

Coefficients

.

Refer to Chapter 5 for Calibration coefficients

.

wu

wu,

State

Wakeup.

Turn ON the analog power (if it is in sleep mode), and send current 

State

 of RINKO

FT. Refer to Chapter 7 for 

State

. See also Chapters 3-a and 8 for 

Error Code

 0003.

querys

querys,

State

Display current state.

Send the current 

State

 of RINKO FT. Refer to Chapter 7 for 

State

.

fwver

fwver=S

Display firmware version.

Send the firmware version of RINKO FT. “S” is a string, and its format is

“Ver.X.XX”.

-

error=

Error Code

Error response

Send the 

Error Code

. Refer to Chapter 8 for 

Error Code

.

Note 1

: RINKO FT returns 

Error Code

 0003 to ANY FIRST COMMAND during the sleep mode. RETRY 

Request

.

Note 2

: Retry 

Request

 if there is something strange, such as no response, error response, or unmatched 

Checksum

.

[Appendix 1] 2 /4

Содержание RINKO FT

Страница 1: ...al DO sensor for integration RINKO FT Model name ARO FT AROD FT User s Manual 1 Use after thoroughly reading this manual 2 Improper use can lead to accidents 3 Safely keep this manual in order not to...

Страница 2: ......

Страница 3: ...ain unit without changing the float inner pressure for calibration inspection Span zero calibration is possible by the calibration kit optional Table of Contents 1 Terminology and Part Names 2 2 Packa...

Страница 4: ...Refers to values such as water temperature and DO calculated from the AD value with calibration coefficients Standard type ARO FT Refers to RINKO FT with standard mounting attachment Cable type ARO F...

Страница 5: ...nting attachment Insulation screw Figure 1 2 Cable type ARO FT Figure 1 1 Standard type ARO FT Main Unit Oxygen detection film Temperature sensor Lock nut Locking sleeve LEMO connector Figure 1 3 MCBH...

Страница 6: ...r 3m cable with LEMO connector and pigtail end 1 4 User s Manual this manual 1 5 Calibration certificate 1 Note the above list indicates the basic package The content may differ depending on the order...

Страница 7: ...1 3 Dummy cap 1 4 User s Manual this manual 1 5 Calibration certificate 1 Note The above list indicates the basic package The content may differ depending on the order 2 Optional parts No Items QTY 1...

Страница 8: ...a hazardous situation which if not avoided could result in minor or moderate injury Indicates general precautions Indicates supplements to control method and convenient functions Danger Ensure safety...

Страница 9: ...sure to insert the connector completely up to its base It may cause seeping of water into the instrument Do not leave the instrument in a place with high temperature e g under direct sunlight for a lo...

Страница 10: ...o the power supply 8 When using the instrument in water it is recommended to use it at 30 C or below When it is used in water at 30 C or above the tracking for correction of temperature change can slo...

Страница 11: ...N C 2 Grey TX N C 2 5 1 2 Cable type ARO FT The Cable type is only for RS 232C communication RX and TX are from the instrument side Green RS 232C Red Power input 6 26 VDC Black GND Green FG Yellow RX...

Страница 12: ...g hold the metal part and pull it out Pulling only the cable can break it Insert the connector into the main unit until two black O rings are completely hidden Hold it in while going to the next step...

Страница 13: ...ut by loosening the locking sleeve Before attaching again apply adequate amount of grease on the O rings When detaching hold the metal part and pull it out Pulling only the cable can break it Insert t...

Страница 14: ...ach It can be pulled out by loosening the locking sleeve Before attaching again apply adequate amount of grease on the O rings When detaching hold the connector and pull it out Pulling only the cable...

Страница 15: ...by water gently remove it with a soft cloth Rubbing it hard can damage the oxygen detection film 2 After water cleaning remove the moisture either by an air blower or by wiping with a soft cloth 3 Ch...

Страница 16: ...ion starts automatically go to 5 If it does not start go to 3 3 Open Computer from the start menu and select the drive with the CD ROM 4 Double click setup exe 5 Screens are displayed in the following...

Страница 17: ...dialog may be displayed during installation for security If UAC dialog appears select Yes to continue the installation UAC dialog might not appear depending on the computer setup but this is not abnor...

Страница 18: ...is connected to the computer 2 The driver needs to be installed if it displays the screen below Follow the steps below Following procedures are not necessary if the driver is properly installed 3 Ope...

Страница 19: ...n below 6 Right click the item with the mark and click Update Driver Software 7 Click Browse my computer for driver software 8 Click Browse button and select the CD ROM or the ARO FT Driver folder ins...

Страница 20: ...his driver software anyway 10 If it displays the screen below you have succeeded with the driver installation Check the Device Manager to make sure that the marks such as and have disappeared If the m...

Страница 21: ...Manager if it is View by Category Right click Start Computer and search Properties Device Manager b Check Port COM and LPT 3 Start the specialized software FT Calib see 7 4 1 4 Select the port searche...

Страница 22: ...7 4 2 Close the software The software can be closed by the following method 1 Close from the File menu The software closes when selecting File Close 2 Close by the button on the title bar The software...

Страница 23: ...rer is recommended Place the DO sensor in bubbled water and wait until the value stabilizes When the value stabilizes click the OK button and obtain the value The value can be obtained as many times a...

Страница 24: ...lm if the value is not calibrated 7 5 1 How to make air bubbling water for SPAN calibration 7 5 2 How to make sodium sulfite solution for zero calibration Make the sodium sulfite solution referred to...

Страница 25: ...rated water This calibration uses air bubbled water as the reference Enter Atmospheric Pressure at the time of calibration RMDO Reference Material DO Under development A reference material can be used...

Страница 26: ...ing the Refresh button Click the Select button to select the file to be updated After clicking the Start button two confirmation screens will appear Click the OK button to start the update Restart the...

Страница 27: ...ion 1 Start the software FT Calib see 7 4 1 2 Select Help Version information The Version information window will be displayed The content and specifications may change for correction and improvement...

Страница 28: ...lose without saving Item Description Log Folder The folder where the logs of calibration and communication are saved Timeout Time length of response to communication commands If time out occurs freque...

Страница 29: ...gs are displayed 7 7 2 Display of precautions for calibration A caution for calibration is displayed when starting the software The display of the caution can be changed by Option see 7 7 The display...

Страница 30: ...he guidance 7 8 2 Uninstall using the Programs and Features From Start menu click Control Panel From the Control Panel window select Programs and Features Select FT Calib from the list by clicking it...

Страница 31: ...rong noise source is not in contact with the main unit or the lead wire Check if the measurement value is obtained after the preheat time 5 seconds Check if there is any large scratch on the oxygen de...

Страница 32: ...Set up the connection port see 7 3 The connection port is wrong If the computer has multiple ports a port which is not connected to the instrument may be set up Confirm the port number see 7 3 Time ou...

Страница 33: ...interval 1 s Pre heat time 5 s Model name ARO FT AROD FT Depth rating 2 000 m depth equivalent 6 700 m depth equivalent Material Housing Titanium grade 2 Insulation screw POM Housing Titanium Ti 6Al...

Страница 34: ...8 1 GB or larger 32 bit 2 GB or larger 64 bit HDD Hard disk with space more than 1 GB About 30 MB is needed for installation Resolution 1024 768 pixels or greater Port USB 1 Other NET Framework 3 5 S...

Страница 35: ...33 Drawing 9 4 Dimensions are in mm Figure 9 1 Standard type ARO FT Figure 9 2 Mounting part of standard type ARO FT Insulation screw and O rings Sensor side Mounting hole Platform side...

Страница 36: ...34 Dimensions are in mm Figure 9 3 Cable type ARO FT Figure 9 4 MCBH type AROD FT...

Страница 37: ...compensated oxygen concentration pc is calculated from the equation below pc 1 Pressure MPa Calibration coefficient The salinity compensated oxygen concentration sc is calculated by multiplying the f...

Страница 38: ...damage due to external causes such as fire earthquake flood lightening other natural disasters public nuisance abnormal voltage corrosive gas organic solvent and chemical solution f Malfunction and da...

Страница 39: ......

Страница 40: ...p mode The checksum is the least significant byte of ones complement of the sum of the values of Request or Response and comma The checksum should be converted to 2 byte ASCII For example In case of a...

Страница 41: ...FF when receiving the command it takes 5 seconds for the DO value to stabilize For that reason DO NOT USE this command during the sleep mode because the RINKO FT again enters sleep mode immediately af...

Страница 42: ...e output value 5 means 50 ms In order to calculate the LED accumulated time in seconds LED accumulated time s decimal expression for LED accumulated time HEX 100 F data format F is a real number F is...

Страница 43: ...ternally calculate the AD values of DO The physical values of DO can be converted from the AD values of DO using the equation provided in the manual section 9 5 First response from the sleep mode RINK...

Страница 44: ...d o n HEX Temp HEX DO Temp 4 unfixed T AD value DO 4 DO AD value 15 16 17 18 19 20 21 22 23 24 25 26 27 28 LED 8 LED accumulated time HEX LED CR LF 0 1 2 3 4 5 6 7 8 9 11 tdona Request t d o n a CR LF...

Страница 45: ...e B Temp 4 unfixed T AD value DO 4 DO AD value 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 Phase B 4 Blue phase AD value HEX Phase R HEX Amplitude B HEX Amplitude R Phase R 4 Red phase AD value Ampli...

Страница 46: ...6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 d3 F d 3 F d3 coefficient CR LF d3 Max unfixed DO d3 coefficient coefficient 16 Uchida et al 2013 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 1...

Страница 47: ...Calibration 10 unfixed Date of T calib date 0 1 2 3 4 5 6 7 23 wu Request w u CR LF 0xE7 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 24 wu State Response w u n o r m a l CR LF State normal 0x32 0 1 2 3 4 5 6 7...

Страница 48: ...responds after error 0003 see Chapters 3 and 8 in appendix 1 Example 1 Sample T and DO at 10 s or more of sampling interval It is much better to power on and off RINKO FT for each measurement to save...

Страница 49: ...y host controller Command qs No response Power OFF RINKO FT End OK Repeat sampling Yes No Example 2 Sample T and DO at 5 s or less of sampling interval In this case RINKO FT is necessary to be continu...

Страница 50: ...Manufacturer 3 48 Takahata cho Nishinomiya Hyogo 663 8202 Japan Phone 81 798 66 1783 Fax 81 798 66 1654 URL http www jfe advantech co jp E mail ocean jfe advantech co jp...

Отзывы: