4. Commands:
HC to RINKO FT RINKO FT to HC Description
Request
Response
Outline
Detail
do
do,HEX
Send DO,
enter sleep mode.
Turn ON the analog power (if it is in sleep mode), send the DO physical data [μ
mol/L], and enter sleep mode. If the RINKO FT analog power is OFF when receiving
the command, it takes 5 seconds for the DO value to stabilize. For that reason, DO
NOT USE this command during the sleep mode because the RINKO FT again enters
sleep mode immediately after sending unstable/unreliable data.
tdo
tdo,HEX,HEX
Send T and DO,
enter sleep mode.
Turn ON the analog power (if it is in sleep mode), send the temperature and DO
physical data [
°
C and μmol/L], and enter sleep mode. If the RINKO FT analog power
is OFF when receiving the command, it takes 5 seconds for the DO value to stabilize.
For that reason, DO NOT USE this command during the sleep mode because the
RINKO FT again enters sleep mode immediately after sending unstable/unreliable
data. The left and right HEXs represent temperature and DO, respectively.
sdo
sdo,HEX
Send DO.
Turn ON the analog power (if it is in sleep mode) and send the DO physical data [μ
mol/L]. If the RINKO FT analog power is OFF when receiving the command, it takes
5 seconds for the DO value to stabilize.
stdo
stdo,HEX,HEX
Send T and DO.
Turn ON the analog power (if it is in sleep mode) and send the temperature and DO
physical data [°C and μmol/L]. If the RINKO FT analog power is OFF when
receiving the command, it takes 5 seconds for the DO value to stabilize. The left and
right HEXs represent temperature and DO, respectively.
tdon
tdon,HEX,HEX,H
EX
Send AD values of T and DO
and LED time,
enter sleep mode.
Turn ON the analog power (if it is in sleep mode), and send the temperature and DO
AD values and LED accumulated time, and enter sleep mode. If the RINKO FT
analog power is OFF when receiving the command, it takes 5 seconds for the DO
value to stabilize. For that reason, DO NOT USE this command during the sleep
mode because the RINKO FT again enters sleep mode immediately after sending
unstable/unreliable data. The first HEX represents the temperature, the second the
DO, and the third the LED accumulated time.
tdona
tdona,HEX,HEX,
HEX,HEX,HEX,
HEX,HEX
Send all parameters as AD
values and LED time,
enter sleep mode.
Turn ON the analog power (if it is in sleep mode), send temperature and DO AD
values, phases of the blue and red lights (AD values), amplitudes of the blue and red
lights (AD values), and LED accumulated time, and then enter sleep mode. If the
RINKO FT analog power is OFF when receiving the command, it takes 5 seconds for
the DO value to stabilize. For that reason, DO NOT USE this command during the
sleep mode because the RINKO FT again enters sleep mode immediately after
sending unstable/unreliable data. The first HEX represents the temperature, the
second the DO, the third the blue phase, the fourth the red phase, the fifths the blue
amplitude, the sixth the red amplitude, and the seventh the LED accumulated time.
stdon
stdon,HEX,HEX,
HEX
Send AD values of T and DO
and LED time.
Turn ON the analog power (if it is in sleep mode) and send the temperature and DO
AD values and LED accumulated time. If the RINKO FT analog power is OFF when
receiving the command, it takes 5 seconds for the DO value to stabilize. The first
HEX represents the temperature, the second the DO, and the third the LED
accumulated time.
stdona
stdona,HEX,HEX,
HEX,HEX,HEX,
HEX,HEX
Send all parameters as AD
values and LED time.
Turn ON the analog power (if it is in sleep mode) and send temperature and DO AD
values, phases of the blue and red lights (AD values), amplitudes of the blue and red
lights (AD values), and LED accumulated time. If the RINKO FT analog power is
OFF when receiving the command, it takes 5 seconds for the DO value to stabilize.
The first HEX represents the temperature, the second the DO, the third the blue
phase, the fourth the red phase, the fifths the blue amplitude, the sixth the red
amplitude, and the seventh the LED accumulated time.
qs
qs,OK
Enter sleep mode.
Enter sleep mode after returning a response. When turning off the power of the
RINKO FT, do so after the RINKO FT is in sleep mode.
*serialnumber
*serialnumber=S Display serial number
The serial number of the RINKO FT. “S” is an ASCII string [0-9a-zA-Z]. The length
of the string is 10 at maximum.
dc
(1)dc,OK
(2)
Coefficients
Display calibration
coefficients.
(1) First, it gives a response to the command (
Response
=dc,OK).
(2) Next, returns the calibration
Coefficients
.
Refer to Chapter 5 for Calibration coefficients
.
wu
wu,
State
Wakeup.
Turn ON the analog power (if it is in sleep mode), and send current
State
of RINKO
FT. Refer to Chapter 7 for
State
. See also Chapters 3-a and 8 for
Error Code
0003.
querys
querys,
State
Display current state.
Send the current
State
of RINKO FT. Refer to Chapter 7 for
State
.
fwver
fwver=S
Display firmware version.
Send the firmware version of RINKO FT. “S” is a string, and its format is
“Ver.X.XX”.
-
error=
Error Code
Error response
Send the
Error Code
. Refer to Chapter 8 for
Error Code
.
Note 1
: RINKO FT returns
Error Code
0003 to ANY FIRST COMMAND during the sleep mode. RETRY
Request
.
Note 2
: Retry
Request
if there is something strange, such as no response, error response, or unmatched
Checksum
.
[Appendix 1] 2 /4