B-88-1006, REV. B 

 
2.0  OPERATION 
 
2.1  Using the Controls; Meaning of the Indicators: 

 

 
  There are three 2-position control switches on the front panel of the SCU (Figure 1). The first 
is the ON-OFF power switch, the second is the HIGH-LOW gain switch, and the third is the 
ON-OFF low-pass filter switch. The ratio between HIGH and LOW gain for your SCU is given in 
Appendix B, as are the filter time constants corresponding to the ON and OFF filter switch 
positions. 
 
  There are two terminal strips on the back panel of the SCU (Figure 1). The lower terminal 
strip provides the connections to two tilt transducers, and one temperature sensor. The upper 
terminal strip provides the input connection for your external power supply, and the connection 
for the conditioned output of the tilt and temperature sensors. The names and functions of the 
terminal strip positions on the rear panel of the SCU, are listed below: 
 
 
 

NAME

 

FUNCTION

   

 

 

 
  SENSOR 1 

The three terminal positions (+), (E), and (-) are for the  

   

three lead wires from the first tilt sensor. (E) is the center, 

   

or neutral wire. (+) and (-) are the positive and negative  

   

pick-ups respectively. For color coding of the sensor lead  

   

wires supplied with your SCU, see Appendix B. 

 
 
  SENSOR 2 

The three terminal positions (+), (E), and (-) are for the  

   

three lead wires from the second tilt sensor. 

 
 
  TEMP. SENSOR 

The three terminal positions (T1), (T2), and (T3) are for  

   

the three lead wires from the LM-35 temperature sensor.   

   

For the color coding of the temperature sensor lead wires  

   

supplied with your SCU, see Appendix B. 

 

 
 
  OUT +1 

This is the Sensor 1 output signal. When operating the  

   

SCU in single-end mode, this signal is referenced to  

   

SIG GND. When operating the SCU in differential mode,  

   

this signal is referenced to OUT -1. 

 
 
  OUT -1 

The signal carried on this terminal position is the OUT +1  

   

signal multiplied by -1.  OUT +1 is referenced to OUT -1  

   

when a differential output is desired. 

 
 
  OUT +2 

This is the Sensor 2 output signal. When operating the  

   

SCU in single-end mode, this signal is referenced to  

   

SIG GND. When operating the SCU in differential mode,  

   

this signal is referenced to OUT -2. 

 

Содержание 781

Страница 1: ...T Serial No ________________ 850 Perimeter Road Manchester NH 03103 USA Tel 603 669 6400 Fax 603 622 2690 sales jewellinstruments com www jewellinstruments com Copyright 2013 by Jewell Instruments LLC All rights reserved Document No B 88 1006 Rev B ...

Страница 2: ... 1 Introduction 2 1 2 General Specifications 2 2 0 Operation 6 2 1 Using the Controls Meaning of the Indicators 6 3 0 Maintenance and Troubleshooting 7 3 1 Routine Maintenance 7 3 2 Determining the Cause of Malfunctions 8 A Warranty and Assistance 9 B Custom Specification for your Equipment 10 C Revision Table 12 ...

Страница 3: ... control switches on the front panel of the SCU and the input and output terminals on the rear panel Figures 1 and 2 Basic maintenance and troubleshooting guidelines are given in the last section of the manual 1 2 General Specifications INPUT CHANNELS Two Jewell Instrument electrolytic tilt sensors one LM 35 temperature sensor TILT OUTPUT Two channels of single ended and differential analog output...

Страница 4: ...l barrier strips TEMPERATURE RANGE 20 C to 70 C 4 F to 158 F Operational 30 C to 100 C 22 F to 212 F Storage HUMIDITY RANGE 0 to 100 noncondensing nonsubmersible SIZE 3 5 x 5 x 8 4 8 8 cm x 12 7 cm x 21 3 cm WEIGHT 2 lbs 0 9 kg MATERIALS Case painted aluminum ...

Страница 5: ...B 88 1006 REV B 4 ...

Страница 6: ...B 88 1006 REV B 5 ...

Страница 7: ...three terminal positions E and are for the three lead wires from the first tilt sensor E is the center or neutral wire and are the positive and negative pick ups respectively For color coding of the sensor lead wires supplied with your SCU see Appendix B SENSOR 2 The three terminal positions E and are for the three lead wires from the second tilt sensor TEMP SENSOR The three terminal positions T1 ...

Страница 8: ... DC to this terminal PWR GND Connect the power ground power common to this terminal 12 V Connect 11 to 15 volts DC to this terminal 3 0 MAINTENANCE AND TROUBLESHOOTING 3 1 Routine Maintenance The SCU is rugged and should require no maintenance other than occasional cleaning of the box with a soft cloth Keep the SCU and tilt sensor from extremes of heat and cold Extreme temperatures unnecessarily s...

Страница 9: ...in each side panel then gently lift away the SCU top panel Small plastic connectors are used to attach the tilt and temperature sensor leads to the printed circuit board Check to be sure that these connectors have not become loose or detached If you remove any of these small plastic connectors be sure to replace the Sensor 1 Sensor 2 and temperature connectors exactly as you found them Figure 2 sh...

Страница 10: ...oid if the equipment is subjected to lighting strikes or other large potential gradients or if it is otherwise abused After expiration of the warranty Jewell will repair the equipment at its factory for parts and labor charges Products returned after warranty expiration should be accompanied by a purchase order to cover repair costs Jewell Instruments is not liable for consequential damages The re...

Страница 11: ...ting is listed below After an instantaneous change in tilt the output signal settles to 90 of this value after three time constants and to 98 of this value after four time constants The corner or cutoff frequency ƒc is defined as the frequency at which signal attenuation is 3 dB Filter roll off above the corner frequency is constant at 6 dB per octave 20 dB per decade Corner frequency can be calcu...

Страница 12: ... Scale Factors are expressed as angular change per volt differential Scale Factors are one half single ended Scale Factors SCU serial number ____________ Calibration range units ____________ Scale Factor units ____________ SENSOR 1 CHANNEL Transducer Serial No Calib Temp F Calib Range Scale Factors Single ended Differential High Gain Low Gain High Gain Max Nonlinearity Low Gain Max Nonlinearity SE...

Страница 13: ...B 88 1006 REV B 12 APPENDIX C Revision Table REV PAGE NOS ECN NO DESCRIPTION OF CHANGE DATE B ALL 25087 Jewell was Applied Geomechanics Added Appendix C for revision record 2 13 13 ...

Отзывы: