Jeulin TECHNOSKLO DP 4000 Скачать руководство пользователя страница 9

Caution 

Close the water tap.

Switch off the apparatus (mains switch), and allow it to cool down for at least 15
minutes.

Clear the defect in the water supply.

Switch  off  the  temperature  limiter  :    take  a  piece  of  non-conducting  material  (length
minimal 50 and maximal  outside diameter 3 mm), insert it to the hole (14) on the back
side  of  the  base  of  the  apparatus  and slightly  press  on it  .  Remove used  piece  of  non-
conducting material from the hole (14).

To continue distilling repeat the procedure starting at 

XI. Putting into operation

.

XIII. To switch off

Press the main switch. The green lamp will go out.

Close the water tap.

Pull out mains supply plug out of the mains socket.

Caution 

The water still must have cooled down completely before the water can be drained through 
the outlet valve (9). Close outlet valve afterwards. 

XIV. Cleaning

Deposits (e.g. lime) build up during the distillation process on the interior parts of evaporator 
(11). This gradually reduces the quality of the distillate. 
When there are considerable deposits in the evaporator, or when the quality of the distillate no 
longer meets the requirements for the intended application, the evaporator must be cleaned. 
When there are heavy deposits in the cooling spiral, they must be cleaned up latter; these 
deposits however do not affect the quality of the distillate. 

Producer cannot accept any warranty claims if the heater is damaged as a result of overheating 
because the apparatus has not been sufficiently cleaned. 

The following items are required for cleaning: 

protective clothing, gloves and goggles

a suitable funnel

a funnel stand, minimal length 70 cm

a graduated jug of minimal capacity 2 litters

a suitable vessel for the disposal of the acetic acid

Содержание TECHNOSKLO DP 4000

Страница 1: ...ratus for producing of distilled water TYPE DP 4000 Operating manual TECHNOSKLO Ltd Držkov 135 468 24 Držkov Czech Republic tel 00420 483 346 403 fax 00420 483 385 040 Skype miroslav bruna e mail sales technosklo com http www technosklo com ...

Страница 2: ...of the Tubing 7 Connecting of the cooling water supply 7 Connecting of the water drainage 7 Connecting of the outlet valve 7 XI Putting into Operation 7 Preparing to operation 7 To adjust the water supply 8 XII Interruption of water supply during working 8 XIII To switch off 9 XIV Cleaning 9 To clean the evaporator 10 To clean the cooling spiral 12 XV Return of the apparatus for the Repair 12 XVI ...

Страница 3: ...er consumption Rated current Electrical protection tube fuse 5 x 20 F 8 A Mains voltage Distilled output Conductivity of distillate referred to 25 C Distilled temperature Cooling water consumption Dimensions of the apparatus W approx A approx A V Hz l h µS cm C l h mm 3100 15 1 230 50 4 1 5 2 5 95 50 190 x 190 x 600 III General instructions and safety regulations 1 To avoid possible hazards pay go...

Страница 4: ... only Do not use old hoses 15 Never send the apparatus to your supplier for repair without having cleaned it first Caution Do not operate the distillation apparatus unless it is properly connected to an earthen electrical installation which is in perfect condition Make sure that earth connection complies with valid electrotechnic regulations Otherwise serious hazard of persons and fire endanger ca...

Страница 5: ...t valve 2 reservoir bottle 10 mains switch 3 tubing connection 11 evaporator 4 distillate nozzle 12 green lamp mains 5 protective cap 13 degassing outlet 6 level tube 14 hole for switch off the temperature limiter 7 tubing connection for cooling water supply 15 cooling coil 8 overflow tubing connection ...

Страница 6: ...f the sink The minimum safety distance from the supply water drain should be 1 m to avoid the risk of splashing Never set up the apparatus closer than this VIII Installing of the Reservoir Bottle In order to achieve conductivity of distillate water mentioned in Technical data apparatus and the Reservoir Bottle must form closed system Carbon dioxide from the air dissolves in water and would cause a...

Страница 7: ...nection 8 to the sink The water level in the evaporator is given by position of the bent tubing connection 8 on the tubing 6 Fit the end of the PVC tubing outer diameter 19 mm supplied to overflow tubing connection 8 The second end lead to the sink and in a case of necessity cut it off to suitable length Insert the second end to the sink and safe the tube against falling out off the sink Don t all...

Страница 8: ...ter supply until it is just sufficient to stop the steam from escaping The water supply is thus optimally adjusted Note When operating the apparatus for the first time let it run distilling in distilling mode for one hour Discard the distillate Attention Do not touch the still body Danger of burns Attention Risk of electric shock Attention Keep of accidental stroke of the hard material to glass pa...

Страница 9: ...ed through the outlet valve 9 Close outlet valve afterwards XIV Cleaning Deposits e g lime build up during the distillation process on the interior parts of evaporator 11 This gradually reduces the quality of the distillate When there are considerable deposits in the evaporator or when the quality of the distillate no longer meets the requirements for the intended application the evaporator must b...

Страница 10: ...e tubing leading from the overflow tubing 8 If necessary dip the end of the tubing into hot water approximately 80 C for about 3 minutes before fitting Check that funnel fits tightly Support the funnel in the stand The bottom of the funnel cone must be level with the distillate drain tube 4 view picture on this page Slowly pour in approximately 1 litter of acetic acid 10 50 using the graduated jug...

Страница 11: ...stillate drain tube 4 drain it through the outlet valve 9 and close the outlet valve again repeat the rinsing three times remove the funnel and hang the overflow tubing into the sink Unscrew the cap 5 and keep it on the base of the apparatus behind the level tube After cleaning Let the apparatus run in distilling mode for one hour Discard the distillate ...

Страница 12: ...el cone must be level with the degassing outlet 13 see fig on this page Slowly pour in approximately 300 ml of acetic acid 10 50 using the graduated jug Be carefully and avoid spilling After cleaning Let the apparatus run in distilling mode for one hour Discard the distillate Caution Acetic acid flows out outlet valve and overflow tubing into the vessel As soon as the deposits have dissolved pour ...

Страница 13: ... recess in the base Put on the both halves of the clamping ring The holes of the halves of clamping ring must be exactly above the holes in the base and second silicone o ring must fit into the groove in the halves of clamping ring Insert screws to the holes of the clamping ring and gradually tighten the screws with using of Allen key Never fully tighten one half of the clamping ring and subsequen...

Отзывы: