JEOL 6060LV Скачать руководство пользователя страница 6

running auto alignment first select an area on the sample or sample holder that has 
homogeneous brightness at moderate magnification (1-5 kx) to get the best beam 
alignment.   

18.

 

Select a spot size, usually initially between 30 and 40.  Click 

 to 

get dialog box with slider control. Usable range is about 10~75. 

 

19.

 

If desired, chose a pre-defined 

Recipe

 to set microscope conditions

 (you may 

switch to LV mode by selecting an LV recipe). This is a very good way to set 
appropriate SEM parameters if you don’t know what you want. 

 

Select the desired recipe and click execute. 

20.

 

Click 

View

 

 {sets fast scan (same as Scan 2) at very low magnification (30X)}. 

At this point, you could also select Scan 2 and move stage to area of interest an select 
a relatively low magnification (features still visible although they may be out of 
focus).

 

21.

 

Click 

ACB

 (auto contrast brightness).

 

22.

 

Focus and check the objective aperture centering

, especially for high 

magnification work.

 

23.

 

Fine focus and correct the astigmatism the image manually

 or click 

Auto Focus

 

 or 

Auto Stigma

 

 (auto focus and auto stigmator). On many samples 

auto focus and auto stigmate actually will fail or will not be exact. Usually you need 
to manually focus and stigmate especially at higher magnifications. 

Almost always 

you need to fine tune focus and stigmators manually!! 

These functions are, of 

course, available on the knob pad.  You also need to check and, if needed, adjust the 

objective aperture centering

. See the next step and individual procedures in the rest 

of the manual for details on how to perform these operations.

 

 
24.

 

Observe your sample  
a.

 

If the image shifts when manually focusing, align the objective aperture. 
Adjusting this will always cause the astigmatism correction to change! 

b.

 

If the image focuses asymmetrically or with stretching or streaks, correct the 
astigmatism. 

Содержание 6060LV

Страница 1: ...computer use the account USER jeol PASSWORD jeol Please do not log into other accounts Use portable media ftp or UIUC NetFiles to transfer your data 3 Prepare your samples before venting Please keep...

Страница 2: ...the height gauge for desired working distance No part of the sample should extend above the ruler end A 10 mm W D should be used for nearly all work except low magnification imaging or if you need to...

Страница 3: ...contact these when observing your sample 10 Turn on the IR chamber scope if off using the switch button on the control box and a switch on CRT monitor and adjust illumination as needed Always use cham...

Страница 4: ...filament heating or Auto Gun Control We found the use of these seem to shorten filament life significantly a Select an area on the sample or sample holder that has reasonably homogeneous brightness a...

Страница 5: ...ases after going through a maximum before beginning of orange indicator the gun tilt shifts are misaligned This is usually because the filament was significantly under saturated when you started Resta...

Страница 6: ...cus 21 Click ACB auto contrast brightness 22 Focus and check the objective aperture centering especially for high magnification work 23 Fine focus and correct the astigmatism the image manually or cli...

Страница 7: ...rging artifacts h Many commonly used functions are available on the knob panel magnification brightness contrast stigmator x and y image fine shift x and y scanning mode and auto functions focus stigm...

Страница 8: ...chassis stops blinking slide open stage drawer 6 If changing samples go to step 7 of general operation 7 Change vacuum mode to HV if in LV mode 8 Carefully slide the stage drawer closed Press EVAC bu...

Страница 9: ...other translation control Iterate to fine tune Correct any severe astigmatism see below to be able to better fine tune 5 Deselect Wobb Manually Correcting the Astigmatism Note You need an image with...

Страница 10: ...10 30 kV is the condition the instrument provides it s best resolution 100 kX possible This is a fairly noisy image even on a Au C sample You must choose the spot size according to your sample and wh...

Страница 11: ...focus astigmatism correction a little more difficult to do ACB and manual knob adjustment function in all the BEIW modes Selecting the Operating Voltage 1 Click Acc Volt to get 2 Double Click value t...

Страница 12: ...samples but this requires addition procedures wet In general use as low as pressure as possible that works for imaging the sample 1 to 10 s of Pa Again BEIW Shadow mode provides secondary electron lik...

Страница 13: ...computer is responding use the software restart START Menu If the computer is locked press and hold the power button on the front of the computer until it turns off Release the button then press it a...

Страница 14: ...d Manual control buttons Manual adjusting buttons spotsize contrast brightness etc and image size switching button are arranged Image display area A SEM image of 640 480 pixels can be displayed and th...

Отзывы: