JEOL 6060LV Скачать руководство пользователя страница 12

Changing the Vacuum Mode 
 
LV mode is generally used to observe and analyze non-conducting sample without 
the need for a conductive coating (Au/Pd or C). 
 

1.

 

Turn off the Beam, i.e. Click 

 

2.

 

Click 

Vac. Mode 

  

A dialog box confirming the change of mode should appear. 

3.

 

Click 

OK

 

4.

 

When switching to LV mode, a Low Vacuum Control Dialog box appears. 

5.

 

An operating pressure of 

1 to 270 Pa

 (2 torr) can be selected. 

6.

 

When 

HT Ready 

 re-appears, click it to restore the e-beam (

LV mode

 can be used to look at and analyze uncoated non-conductive samples 

without charging artifact and imaging problems.  It can also be used to observe wet or 
badly out-gassing samples, but this requires addition procedures (wet).  In general, 
use as low as pressure as possible that works for imaging the sample (1 to 10’s of Pa).  
Again 

BEIW

-

Shadow

 mode provides secondary electron like images. Since the e-

beam will interact with the gas in the specimen chamber, increasing pressure will 
result in a signal decrease and the resolution will get poorer.  Also, longer working 
distances generally will not work well in 

LV mode

 
Starting with LV vacuum mode (sample is not suitable for HV vacuum mode) 

Change mode to LV before pressing EVAC. Automatic gun alignment functions do not 
function in LV mode. Strongly suggest verifying proper operation/alignment in normal 
HV mode first (even if it is just on the stage, i.e. no sample) 

 
Improving Depth of Focus or Obtaining Lower Magnifications 

Move sample to a longer working distance. 
 

 

 

Содержание 6060LV

Страница 1: ...computer use the account USER jeol PASSWORD jeol Please do not log into other accounts Use portable media ftp or UIUC NetFiles to transfer your data 3 Prepare your samples before venting Please keep...

Страница 2: ...the height gauge for desired working distance No part of the sample should extend above the ruler end A 10 mm W D should be used for nearly all work except low magnification imaging or if you need to...

Страница 3: ...contact these when observing your sample 10 Turn on the IR chamber scope if off using the switch button on the control box and a switch on CRT monitor and adjust illumination as needed Always use cham...

Страница 4: ...filament heating or Auto Gun Control We found the use of these seem to shorten filament life significantly a Select an area on the sample or sample holder that has reasonably homogeneous brightness a...

Страница 5: ...ases after going through a maximum before beginning of orange indicator the gun tilt shifts are misaligned This is usually because the filament was significantly under saturated when you started Resta...

Страница 6: ...cus 21 Click ACB auto contrast brightness 22 Focus and check the objective aperture centering especially for high magnification work 23 Fine focus and correct the astigmatism the image manually or cli...

Страница 7: ...rging artifacts h Many commonly used functions are available on the knob panel magnification brightness contrast stigmator x and y image fine shift x and y scanning mode and auto functions focus stigm...

Страница 8: ...chassis stops blinking slide open stage drawer 6 If changing samples go to step 7 of general operation 7 Change vacuum mode to HV if in LV mode 8 Carefully slide the stage drawer closed Press EVAC bu...

Страница 9: ...other translation control Iterate to fine tune Correct any severe astigmatism see below to be able to better fine tune 5 Deselect Wobb Manually Correcting the Astigmatism Note You need an image with...

Страница 10: ...10 30 kV is the condition the instrument provides it s best resolution 100 kX possible This is a fairly noisy image even on a Au C sample You must choose the spot size according to your sample and wh...

Страница 11: ...focus astigmatism correction a little more difficult to do ACB and manual knob adjustment function in all the BEIW modes Selecting the Operating Voltage 1 Click Acc Volt to get 2 Double Click value t...

Страница 12: ...samples but this requires addition procedures wet In general use as low as pressure as possible that works for imaging the sample 1 to 10 s of Pa Again BEIW Shadow mode provides secondary electron lik...

Страница 13: ...computer is responding use the software restart START Menu If the computer is locked press and hold the power button on the front of the computer until it turns off Release the button then press it a...

Страница 14: ...d Manual control buttons Manual adjusting buttons spotsize contrast brightness etc and image size switching button are arranged Image display area A SEM image of 640 480 pixels can be displayed and th...

Отзывы: