JEOL 6060LV Скачать руководство пользователя страница 3

9.

 

Specimen tilting is limited by working distance and specimen diameter.   

The IR 

chamber-scope must be used for sample tilting and Z adjustment to verify that safe 
clearance with detectors, etc. is maintained.

  Safe clearance is at least 5mm distance 

on LCD monitor. The following table shows the typical Allowable Tilt Range for 
samples up to 32 mm diameter. Larger samples are more restricted. This assumes no 
part of sample is higher than indicated Working Distance. 
 

Working Distance 

Allowable Tilt Range 

10 mm 

0 to 15 degrees 

15 mm 

-10 to 15 degrees 

20 mm 

-10 to 35 degrees 

40 mm 

-10 to 65 degrees 

48 mm (max W.D.) 

-10 to 70 degrees 

  

 

Note: The Backscatter detector and SE detector are very delicate and will be 
damaged by any contact with a sample (>$3000 to repair). You must be extremely 
careful not to contact these when observing your sample!!!  

 

 
10.

 

Turn on the IR chamber-scope, if off, using the switch (button) on the control box and 
a switch on CRT monitor and adjust illumination as needed. 

Always use chamber-

scope for tilt and Z adjustment of sample during operation. Note: The chamber-
scope illumination does not interfere with SEI but must be turned off for BEIW 
modes and EDS operation. 

 

11.

 

Carefully (slowly) slide the stage drawer closed, while verifying safe clearance of 
sample on chamber-scope.

  Press 

EVAC

 button on the Specimen exchange dialog 

box. Hold the door closed until pumping starts (it takes a few seconds). 

12.

 

In about 1.5 minutes, the HT icon turns to 

HT ready

   

Wait at least 2 more additional minutes!!

 

13.

 

Select the desired accelerating voltage (up to 30 kV) 

Double click on new value to change setting. 

14.

 

Set the working distance (similar to previous step). Just select the working distance 
you preset the sample for (usually 10 mm). 

SE 
detector 

Conical 
objective  lens

Backscatter 
detector 

Содержание 6060LV

Страница 1: ...computer use the account USER jeol PASSWORD jeol Please do not log into other accounts Use portable media ftp or UIUC NetFiles to transfer your data 3 Prepare your samples before venting Please keep...

Страница 2: ...the height gauge for desired working distance No part of the sample should extend above the ruler end A 10 mm W D should be used for nearly all work except low magnification imaging or if you need to...

Страница 3: ...contact these when observing your sample 10 Turn on the IR chamber scope if off using the switch button on the control box and a switch on CRT monitor and adjust illumination as needed Always use cham...

Страница 4: ...filament heating or Auto Gun Control We found the use of these seem to shorten filament life significantly a Select an area on the sample or sample holder that has reasonably homogeneous brightness a...

Страница 5: ...ases after going through a maximum before beginning of orange indicator the gun tilt shifts are misaligned This is usually because the filament was significantly under saturated when you started Resta...

Страница 6: ...cus 21 Click ACB auto contrast brightness 22 Focus and check the objective aperture centering especially for high magnification work 23 Fine focus and correct the astigmatism the image manually or cli...

Страница 7: ...rging artifacts h Many commonly used functions are available on the knob panel magnification brightness contrast stigmator x and y image fine shift x and y scanning mode and auto functions focus stigm...

Страница 8: ...chassis stops blinking slide open stage drawer 6 If changing samples go to step 7 of general operation 7 Change vacuum mode to HV if in LV mode 8 Carefully slide the stage drawer closed Press EVAC bu...

Страница 9: ...other translation control Iterate to fine tune Correct any severe astigmatism see below to be able to better fine tune 5 Deselect Wobb Manually Correcting the Astigmatism Note You need an image with...

Страница 10: ...10 30 kV is the condition the instrument provides it s best resolution 100 kX possible This is a fairly noisy image even on a Au C sample You must choose the spot size according to your sample and wh...

Страница 11: ...focus astigmatism correction a little more difficult to do ACB and manual knob adjustment function in all the BEIW modes Selecting the Operating Voltage 1 Click Acc Volt to get 2 Double Click value t...

Страница 12: ...samples but this requires addition procedures wet In general use as low as pressure as possible that works for imaging the sample 1 to 10 s of Pa Again BEIW Shadow mode provides secondary electron lik...

Страница 13: ...computer is responding use the software restart START Menu If the computer is locked press and hold the power button on the front of the computer until it turns off Release the button then press it a...

Страница 14: ...d Manual control buttons Manual adjusting buttons spotsize contrast brightness etc and image size switching button are arranged Image display area A SEM image of 640 480 pixels can be displayed and th...

Отзывы: