Err 5
Light Saturation Not Found
This error indicates that the peak light hasn’t been found at zero.
The most likely causes of this error are:
Fatal
1.
Lamp failure.
2.
Deterioating lamp signal.
3.
Sample or cuvette in the sample holder.
Solution
: Ensure that the sample holder is empty. Restart the unit, if the
problem persists contact a service technican.
Err 7
USB Module Failure
This error indicates the instrument is unable to detect a response from
the UBS module.
Warning
The most likely causes of this error are:
1.
USB port failure.
Solution:
press the key adjacent to the
tick
icon to carry on using the
instrument. Restart the unit, if the problem persists contact a service
technician.
Err 8
Unable to find Vane on Turret
This error indicates that the vane position zero on the turret cannot
be found.
Warning
The most likely causes for this error are:
1.
The turret carousel has been removed and not replaced.
Redo
Solution
: Check if the turret is in the sample chamber and inserted
properly. Press the key adjacent to the
back
icon to try rechecking.
Err 9
Over Temperature
This error indicates that the thermal switch has cut out.
The most likely causes for this error are:
Fatal
1.
Failure of the fan
2
. Thermal switch not connected.
Solution
: restart the unit, if the problem persists contact a service
technician.