SECTION 13 – Troubleshooting
13.1
ERROR CODES
If an error code is displayed it will be accompanied by a
spanner
icon and a symbol to indicate if the
error is a warning (
caution
icon) or fatal (
stop
icon). If the error is fatal contact your local distributor
or Jenway service department (refer to section 12.3). If the error is a warning it may be possible to
retry the test. In this case a
back
icon will also be displayed. The table below shows the error codes:
Error Code Symbol
Issue
Err 1
System Parameters Failure
This error indicates that the essential system parameters are corrupt.
The most likely causes of this error are:
1.
Fram chip failure.
Solution
: Restart the unit, if the problem persist, contact a service
technician.
Err 2
Operating Parameters Failure
This error is a warning that the method parameters have been reset.
The most likely causes of this error are:
Warning
1.
The methods were reset by pressing the top left button during the
power on tests performed when the instrument is turned on.
2.
the parameters have been corrupted, so the instrument has reset
itself.
Solution
: Press the key adjacent to the
tick
icon to action.
Err 3
Dark Calibration Error
This error indicates that the dark level is too high during calibration. In
normal operation the lamp is switched off during an operator initiated
Warning
calibration sequence to ensure that the detector output is below a
threshold level. The calibration is aborted and Err 3 is displayed if the
detector output is above the threshold level.
The most likely causes of this error are:
1.
The sample chamber lid has been left open during the calibration
sequence.
Redo
2.
The sample chamber lid was opened during the calibration sequence.
3.
There is a fault on the detector PCB.
Solution
: Ensure the sample chamber is closed completely. Press the key
adjacent to the
back
icon to retry the dark calibration.
Err 4
Microswitch Failure
This error indicates that the microswitch was not found.
The most likely causes of this error are:
Fatal
1.
The microswitch is broken.
73