jenway 7315 Скачать руководство пользователя страница 52

54

In the expanded operating menu press the key adjacent 
to  the 

USB  memory  stick

  icon  to  open  the  results 

selection menu.

If the USB memory stick doesn’t have any results stored 
on it only the 

save to USB memory stick

 icon will be 

displayed.  To save a result press the key below the 

save 

to  USB  memory  stick

  icon.  This  will  open  the  results 

naming menu.

The  default  result  name  will  be  displayed.  To  save  the 
result under the default name press the key adjacent to 
the 

tick

  icon.  To  change  the  name  of  the  result  press 

the  key  below  the 

eraser

  icon.  One  press  of  the  key 

will delete one letter, holding the key for 2 seconds will 
delete the name completely.

To select letters use the keys adjacent to the 

arrow

 icons to move around the menu. Once the required 

letter is highlighted press the key below the 

pencil

 icon to select the letter. Up to eight characters can 

be selected. To use numbers in the results name press the key below the 

AB

 icon. Repeat pressing of this 

key will cycle between upper case and numbers. Once the required name for the result has been entered 
press the key adjacent to the 

tick

 icon to save and return to the results selection menu.

The saved results are listed alphabetically with the date 
and time that the result was generated.   

If  there  is  already  a  result  saved  under  the  same  name 
confirmation will be needed to replace the existing result 
with the new one. Press the key adjacent to the 

tick

 icon 

to  confirm  replacement.  Press  the  key  adjacent  to  the 

cross

  icon  to  cancel  and  return  to  the  results  naming 

menu. 

In photometrics, concentration and quantitation once a result has been saved to the USB memory stick 
the 

USB memory stick

 icon will remain highlighted. Pressing the key adjacent to the highlighted icon 

will result in subsequent results being saved under the same file name. To save the result with a new 
filename hold the key adjacent to the highlighted

 USB memory stick

 icon for 2 seconds. This will open 

the results naming screen.  

10.5 

OPENING RESULTS

Results can only be opened if a valid USB memory stick is inserted into the front of the instrument. 

In the expanded operating menu press the key adjacent to 
the 

USB memory stick

 icon to open the results selection 

menu.  Select  the  result  to  be  opened  by  pressing  the 
key adjacent to the result.  Press the key below the 

load 

from USB memory stick

 icon.

1

/

1

1.000

0.099

400

09:02

ppm

ABS

nm

JW0000

~

0

A

  B  C  D  E  F  G  H  I 

J  K  L  M  N  O  P  Q  R 
S  T  U  V  W X  Y  Z  _

1

/

1

SAMPLE_1

05/07/2010  11.11

1

/

1

BEER

05/07/2010  11.16

SAMPLE_1

30/07/2010  10.02

Содержание 7315

Страница 1: ...Spectrophotometers Models 7310 7315 Operating Manual 731 005 REV D 06 10...

Страница 2: ......

Страница 3: ...those items defined in the maintenance procedures herein there are no user serviceable items in this instrument Removal of covers and attempted adjustment or service by unqualified personnel will inva...

Страница 4: ...etup 19 4 2 Time and date 20 4 3 Instrument settings menu 20 4 4 Security and setting passwords 21 4 4 1 Setting security codes 21 4 4 2 Settings lock 21 4 4 3 Method lock 21 4 5 Mode selection 22 4 6...

Страница 5: ...ints analysis 36 SECTION 8 Quantitation 38 8 1 Mode specific parameters 38 8 2 Method setup 39 8 2 1 Selecting a wavelength 39 8 2 2 Quantitation table 39 8 2 2 1 Selecting number of standards 39 8 2...

Страница 6: ...results 57 10 8 Autologging 58 10 8 1 Setting the number of sample repetitions 58 10 8 2 Selecting result s destination 59 10 9 Connecting to a PC 59 SECTION 11 Accessories and spare parts 60 11 1 Op...

Страница 7: ...dule replacement 71 12 3 Service 72 SECTION 13 Troubleshooting 73 13 1 Error codes 73 13 2 Troubleshooting guide 75 13 3 Technical support 75 SECTION 14 Declaration of conformity 76 SECTION 15 Glossar...

Страница 8: ...ccuracy 1 T 0 01Abs at 1 000 Absorbance Resolution 0 1 T 0 001A Stray light 0 5 at 340nm 0 5 at 340nm and 220nm Stability 0 002Abs hr after 30 minute warm up 0 001Abs hr without warm up Concentration...

Страница 9: ...Light source Tungsten halogen lamp Xenon lamp Lamp save Yes Not applicable GLP Current time and date user ID settings lock and method lock Number of users 999 Methods memory 48 in each measurement mod...

Страница 10: ...MANCD 8 Optional accessories as ordered 2 2 INSTALLATION Models 7310 and 7315 are supplied ready to use The unit should be placed on a clean flat surface which is free from drafts and vibrations The u...

Страница 11: ...k port and the instrument 2 3 DISPLAY These spectrophotometers have a dot matrix display which enables icons and graphs to be displayed clearly Following successful completion of the power on tests th...

Страница 12: ...above or adjacent to the key The only exception to this is the back key which is always active The main menu screen and surrounding keypad is displayed below Fig 2 4 1 Display 1 Spectrum measurement m...

Страница 13: ...tch for the unit 3 Power in socket Connection socket for power supply unit 4 RS232 serial port Connection to a PC or external serial printer 5 Output sockets Analogue output 2 6 FRONT PANEL The image...

Страница 14: ...ration mol l 1 l is the path length cm This law shows that absorbance is linear to concentration but this is only true for low concentrations For absorbance levels above 3 the concentration starts to...

Страница 15: ...temperature and light levels should remain as constant as possible 2 If required adherence to Standard Operating Procedures SOP and Good Laboratory Practice GLP should be monitored with regular calib...

Страница 16: ...measurement unless a suitable temperature controlled sample holder is in use Temperature change during measurement may cause air bubbles to form on the walls of the sample holder This is a common cau...

Страница 17: ...ast and lamp save menus All of the measurement modes open initially into a minimal operating menu This menu allows calibration and simple readings to be taken without changing any measurement paramete...

Страница 18: ...ease or decrease the number The clock function uses a 24 hour format In the time and date menu to set the date press the key adjacent to the calendar icon Select the digit to be changed using the keys...

Страница 19: ...nction is accessed through the instrument settings menu by pressing the key adjacent to the open padlock icon One press will lock the settings instantly To unlock the settings press the key again This...

Страница 20: ...ment mode icons which are displayed on the main menu are identified with a mode shown icon The mode icons which are not displayed on the main menu are identified with a mode not shown icon To change a...

Страница 21: ...rument settings menu 4 9 LAMP SAVE This function is only available on the 7310 visible spectrophotometer which uses a tungsten halogen lamp The lamp save function enables the time in minutes to be set...

Страница 22: ...n and the lamp will begin to warm up The lamp cold icon is displayed adjacent to the calibrate to zero icon in the measurement mode The time needed for the lamp to warm up is five minutes To view the...

Страница 23: ...ng menu enables measurement parameters to be changed The utility toolbar on the left hand side of the screen enables access to printing print setup options results methods and autologging options For...

Страница 24: ...00 transmittance Once the calibration is complete the measure sample icon appears and the sample can be measured If the wavelength is adjusted before a sample is measured the measure sample icon will...

Страница 25: ...lity toolbar on the left hand side of the screen enables access to printing print setup options results methods and autologging options For more details on the different functions of the utility toolb...

Страница 26: ...as selecting this option will calculate the factor If the factor is known it is not necessary to measure a known standard s concentration When the standard or factor is not selected the value should b...

Страница 27: ...the screen to select the digit to be changed The key below the digit must be pressed twice to select the adjacent digit For example 00 the first press of the key alters 10 the second press alters 01...

Страница 28: ...calibrated at the selected wavelength In this operating mode the type of sample measurement performed depends on the calibration which has been carried out 6 4 1 Measuring a Sample After Calibrating...

Страница 29: ...inting print setup options results methods and autologging options For more details on the different functions of the utility toolbar refer to section 10 The scan settings icon enables the graph y axi...

Страница 30: ...ge from 198 to 1000nm To set the start wavelength in the scan settings menu press the key adjacent to the wavelength on the far left of the x axis This opens the number entry screen Use the keys at th...

Страница 31: ...cted for a wavelength range of 400 to 503nm 7 2 1 4 Y Axis Scaling This function enables the scale of the y axis of the spectrum graph to be adjusted either manually or automatically The automatic sca...

Страница 32: ...l be needed to stop the baseline scan Press the key adjacent to the tick icon to confirm stopping the baseline scan and return to the expanded operating menu Press the key adjacent to the cross icon t...

Страница 33: ...shold value The spectral points analysis function enables points to be selected from the scan to analyse absorbance or transmittance at selected wavelengths 7 5 1 Peaks and Valleys Threshold This func...

Страница 34: ...se the absorbance or transmittance at a selected wavelength To access this function press the key adjacent to the spectral points analysis icon and this opens the selection screen A solid vertical lin...

Страница 35: ...nts analysis table If a wavelength entered is outside the scan range the check number warning icon will be displayed before the wavelength is automatically adjusted to within the range To view the poi...

Страница 36: ...he expanded operating menu The quantitation expanded operating menu enables measurement parameters to be changed The utility toolbar on the left hand side of the screen enables access to printing prin...

Страница 37: ...an also be accessed from this screen by pressing the key adjacent to the standard curve icon The standard curve can also be viewed from this screen by pressing the key adjacent to the standard curve i...

Страница 38: ...on to save the changes The photometric values of the standards can also be entered if known To enter the photometric value press the key adjacent to the photometric value in the table that corresponds...

Страница 39: ...correlation coefficient r2 To exit this screen press the key adjacent to the tick icon The quantitation table can also be viewed from this screen by pressing the key adjacent to the quantitation table...

Страница 40: ...rating menu where unknown samples can be measured When this operating mode is exited the current standard curve will be saved until it is replaced by a new curve The automatic 8 cell turret can be use...

Страница 41: ...istics function is accessed by pressing the key below the S icon in the quantitation curve menu The algorithm of the curve fit can be changed by pressing the key below the curve fit icon For more info...

Страница 42: ...y toolbar on the left hand side of the screen enables access to printing print setup options results methods and autologging options For more details on the different functions of the utility toolbar...

Страница 43: ...h a cross through it and the manual scaling by the hand icon without the cross To select manual or automatic scaling press the key adjacent to the hand icon Repeat pressing of the key cycles between m...

Страница 44: ...far left side of the x axis This opens the lag time start on level menu To set the lag time press the key adjacent to the clock icon Use the keys at the bottom of the screen to select the digit to be...

Страница 45: ...an be reset to zero by pressing the key adjacent to the 000 icon Once the standard value has been entered press the key adjacent to the tick icon to save and return to the settings menu screen The ent...

Страница 46: ...ion must be performed at the same wavelength at which the sample will be measured Insert a cuvette containing the blank solution into the sample chamber and close the instrument lid Press the key belo...

Страница 47: ...nt press the key adjacent to the S icon which opens the statistics menu over the top of the expanded operating menu The information displayed includes concentration absorbance per minute rate of chang...

Страница 48: ...intervals The double greater than or double less than icons move the line in 5 second time intervals Once the required initial and final times have been selected press the key adjacent to the S icon t...

Страница 49: ...the name of the method press the key below the eraser icon One press of the key will delete one letter holding the key for 2 seconds will delete the name completely To select letters use the keys adja...

Страница 50: ...ss the key adjacent to the cross icon to cancel and return to the option of saving one or all of the methods to USB memory stick 10 2 OPENING METHODS Methods can be opened from the internal memory of...

Страница 51: ...to the option of opening one or all methods from the USB memory stick If a single method is selected to be opened this method will be stored in the same location that is was saved in when it was save...

Страница 52: ...aved results are listed alphabetically with the date and time that the result was generated If there is already a result saved under the same name confirmation will be needed to replace the existing r...

Страница 53: ...y adjacent to the result Press the key below the delete icon Confirmation will be needed to delete the selected result press the key adjacent to the tick icon to confirm deletion Press the key adjacen...

Страница 54: ...cted press the key adjacent to the tick icon to save and return to the expanded operating menu 10 7 1 1 Print Setup PHOTOMETRICS AND CONCENTRATION In the photometrics and concentration measurement mod...

Страница 55: ...een a tick and a cross icon for selected or deselected To select the data interval press the key adjacent to the data interval icon Repeat presses will cycle between 0 1 5 10 30 or 60 seconds Once the...

Страница 56: ...ame press the key adjacent to the tick icon To change the name press the key below the eraser icon One press of the key will delete one letter holding the key for 2 seconds will delete the name comple...

Страница 57: ...be recorded Press the key adjacent to the tick icon to confirm selection Press the key adjacent to the cross icon to cancel and return to the expanded operating menu 10 8 2 Selecting Result s Destina...

Страница 58: ...NG THE ACCESSORIES There are two types of accessories which can be fitted in the sample chamber passive or active accessories The range of passive accessories includes 10 x 10mm single cuvette holders...

Страница 59: ...fed correctly 11 2 2 PASSIVE ACCESSORIES Unscrew the thumb screw to undo the passive accessory Lift out the passive accessory To fit a different passive accessory simply place the accessory in the cor...

Страница 60: ...the correct position If the fitting is too tight use a small screw driver to loosen the ball bearings before pushing the carousel down onto the shaft 11 2 3 2 Peltier For this accessory as well as rem...

Страница 61: ...ng must be kept as short as possible and the tubing must not be allowed to obstruct the ligth path For sipping 1 Connect the sipper pump tubing to the outlet port on the flow through cuvette 2 Secure...

Страница 62: ...e motor 3 Fit the other end onto the inlet port on the flow through cuvette 4 Fit the second 300mm length of tubing to the outlet port of the flow through cuvette Once secured ensure the tubing is rou...

Страница 63: ...measured To adjust the wavelength press the key below the 8 cell turret icon and use the arrow icons to adjust the wavelength 11 3 1 1 Automatic 8 cell turret supporting creation of a standard curve...

Страница 64: ...he expanded operating menu The peltier will begin to heat or cool depending on the current temperature 11 3 3 Sipper pump When the sipper is in use the sipper pump icon is displayed in the bottom righ...

Страница 65: ...y below the calibrate timed sipper icon Select the required pump direction by pressing the key below the forwards or backwards arrow icon Press the key adjacent to the tick icon to continue to the nex...

Страница 66: ...r uptake is recorded To fine tune the amount of air uptake press the keys below the plus or minus icons to increase or decrease the amount of air taken in The recorded time will be adjusted accordingl...

Страница 67: ...p take up time If an air gap of zero was previously selected this screen will not appear and the calibration sequence will continue to sample disposal Once this stage of the calibration sequence is co...

Страница 68: ...ier settings enabling the temperature to be set Refer to section 11 3 2 for more details The sipper pump can operate in a manual or timed mode Refer to section 11 3 for more details 11 4 SPARES Part C...

Страница 69: ...panel located on the rear of the instrument refer to section 2 5 1 Remove the screws holding the lamp access panel in place 2 Withdraw the lamp access panel and unscrew the lamp bracket fixing screw...

Страница 70: ...Department for a reference number which you should include with your faulty equipment Please also ensure you include a clear description of the fault and a completed copy of our Decontamination Certif...

Страница 71: ...the top left button during the power on tests performed when the instrument is turned on 2 the parameters have been corrupted so the instrument has reset itself Solution Press the key adjacent to the...

Страница 72: ...adjacent to the tick icon to carry on using the instrument Restart the unit if the problem persists contact a service technician Err 8 Unable to find Vane on Turret This error indicates that the vane...

Страница 73: ...USB memory stick Ensure the USB memory stick has free capacity Ensure there is a result displayed on the screen Unable to continue kinetics Due to the nature of the kinetics measurements stopping the...

Страница 74: ...SECTION 14 Declaration of Conformity 76...

Страница 75: ...SECTION 14 Declaration of Conformity 77...

Страница 76: ...English English icon Language selection Common USB memory stick Common Save to USB memory stick Common Load from USB memory stick Common Result already saved to USB memory stick Common 1 Single metho...

Страница 77: ...screen Common Resolution Common 400nm Wavelength Common ABS T Abs T icon Operating mode either absorbance or transmittance Common Lamp cold Common Lamp save Common Caution icon accompanied by error c...

Страница 78: ...terval between each sample measurement repetition Autolog Batch ABC icon Set batch name for repeated measurements Autolog Batch append add existing results Autolog Batch delete overwrite existing resu...

Страница 79: ...s standard or factor value to zero Concentration Check standard Spectrum Scan settings menu Spectrum Hand icon Manual y axis scaling Spectrum Hand icon with cross through it Auto y axis scaling Spectr...

Страница 80: ...rd curve Quantitation Curve fit enables the curve fit to be adjusted Quantitation Sigma icon Displays curve statistics Quantitation Create new standard curve Quantitation Tick icon Read Standard Quant...

Страница 81: ...single cell Method created on a unit with a single cell accessory fitted Accessories 8 cell turret icon the automatic 8 cell turret is in use Accessories Error finding cell position 0 on turret Acces...

Страница 82: ...s Increase sample uptake increase air gap Accessories No sipper Method created on a unit with a sipper accessory fitted Accessories Sipper peltier pump in use Accessories No sipper peltier Method crea...

Страница 83: ...setup 19 Contrast 23 Expanded operating menu 19 GLP settings 22 Lamp save 23 Mode selection 22 Navigating 19 Utility toolbar 20 KINETICS 44 CALIBRATION 48 DATA ANALYSIS 49 METHOD SET UP 45 MODE SPECI...

Страница 84: ...51 SAVING RESULTS 53 Screen setup 19 SERVICE 71 Spare parts 70 SPECTRUM 31 CALIBRATION 34 DATA ANALYSIS 35 METHOD SETUP 32 MODE SPECIFIC PARAMETERS 31 SAMPLE MEASUREMENT 34 Technical specification 8...

Страница 85: ......

Страница 86: ...Tel 33 1 64 45 13 13 Fax 33 1 64 45 13 00 e mail bsf bibby scientific fr www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Te...

Отзывы: