background image

SECTION 6

INTERFACING

Serial Interface

The Model 6300 has a bi-directional RS232 interface set to:

1200 baud
7 data bits
odd parity
1 stop bit

The 25 way D connector allows a standard one-to-one interconnection lead to be used, as supplied with
the 40 column printer.
A printout is initiated by pressing the PRINT key. If the sample munber is unity, then the printout will
include a header block. The sample number is incremented every time the PRINT key is pressed.

The following commands can also be sent to the 6300 via the serial interface (using Windows Terminal or
Jensoft, for example).

ASCII

 D

 or

 d

Same as pressing the PRINT key

ASCII 

T<CR>

Outputs transmission and wavelength separated by an ASCII TAB character,
regardless of the 6300 operating mode. For example:    100.0   540

ASCII 

A<CR>

Outputs absorbance and wavelength separated by an ASCII TAB character,
regardless of the 6300 operating mode. For example:    0.001   540

ASCII 

C<CR>

Outputs concentration and wavelength separated by an ASCII TAB character,
regardless of the 6300 operating mode. For example:    123.4   540

ASCII 

V<CR>

Outputs a voltage proportional to the monochromatic light level passing
through the sample and wavelength separated by an ASCII TAB character.
For example:

1234.5

540

ASCII 

Z<CR>

Calibrates a zero absorbance if the dark shutter is open (SO<CR> command),
or zero transmittance if the dark shutter is closed (SC<CR> command).

ASCII 

SC<CR>

Closes the dark shutter which blocks monochromatic light entering the
sample chamber. This allows 0% transmittance to be calibrated.

ASCII 

SO<CR>

Opens the dark shutter which allows monochromatic light to enter the sample
chamber. This allows 100% transmittance (zero absorbance) to be calibrated.
The shutter must be open for normal measurements.

ASCII 

Gnnn<CR>

Commands the 6300 to go to the wavelength 

nnm

. For example: G540<CR>

will set the wavelength to 540nm.

6300/REV B/12-96

12

Содержание 6300

Страница 1: ...MODEL 6300 VISIBLE RANGE SPECTROPHOTOMETER OPERATING MANUAL 630 026...

Страница 2: ...tem of work in addition to the detailed instructions specified in this manual 3 The covers on the unit should only be removed by personnel who have been trained to avoid the risk of shock 4 References...

Страница 3: ...N 2 INSTALLATION Unpacking 2 1 Installation 2 2 Displays 2 3 Controls 2 4 Outputs 2 5 SECTION 3 OPERATION Initial Set Up 3 1 Sample Measurement 3 2 Good Practice Guidelines 3 3 SECTION 4 MAINTENANCE G...

Страница 4: ...l contamination Combined with a mechanically rigid structure the 6300 provides a system with fast warm up low drift and high reliability 1 2 INSTRUMENT SPECIFICATIONS Wavelength Range 320 950nm Resolu...

Страница 5: ...your local Distributor before proceeding further Fuse Rating 2A F fast blow type NOTE The unit should be positioned within 1 5 metres of an earthed mains supply VOLTAGE SELECT NOTE When changing the...

Страница 6: ...e Concentration 2 Primary display adjust annunciator 3 Secondary display area Wavelength Factor 4 Primary display units 5 Secondary display adjust annunciator 6 Operation with PC 7 Menu options T ABS...

Страница 7: ...e displayed menu option 4 initiates a calibration routine 5 Print key Provides a printout of the current reading with an incremental sample number When pressed for the first time after a calibration t...

Страница 8: ...mp when replacement is necessary NOTE The Lamp Access Panel and all ventilation slots must not be covered or obstructed at any time 2 ROCKER SWITCH On Off switch for the unit 3 POWER IN SOCKET IEC typ...

Страница 9: ...ot be found the most likely cause is that the sample chamber lid is not closed 3 2 SAMPLE MEASUREMENT Many methodologies are well documented and where these are available the recommended wavelength sh...

Страница 10: ...ctivated to perform the zero setting and this part of the routine is therefore independent of the solution in the light path The following error codes are possible after a calibration Err 1 Dark cal e...

Страница 11: ...ion to the set value by calculating a multiplication factor for the current absorbance If the factor calculated is out of range then the instrument will display Err 3 for 2 seconds after the calibrati...

Страница 12: ...ed surfaces 7 Chemical reagents should wherever possible be of high grade quality Contamination can cause problems even at very low levels Diluents i e water or solvents must be free from impurities 8...

Страница 13: ...urce if this fails Failure should be suspected if lamp failure indicator appears on the display This can be confirmed by looking in the sample chamber The Tungsten Halogen lamp is available from the M...

Страница 14: ...CESSORIES 5 1 OPTIONAL ACCESSORIES The following list of items are available as optional accessories for use with the Model 6300 630 005 Adjustable Cell Holder 630 020 Test Tube Holder 12mm 630 028 Du...

Страница 15: ...ASCII TAB character regardless of the 6300 operating mode For example 0 001 540 ASCII C CR Outputs concentration and wavelength separated by an ASCII TAB character regardless of the 6300 operating mod...

Страница 16: ...active GND 7 Suggested interconnections are detailed below 6300 IBM PC XT 25 way D TXD 2 2 TXD From PC RXD 3 3 RXD To PC RTS 4 4 RTS From PC CTS 5 5 CTS To PC DSR 6 6 DSR To PC DCD 8 8 DCD To PC DTR 2...

Страница 17: ...992 Electromagnetic compatibility Generic immunity standard Performance criterion B EN 61010 1 1993 Safety requirements for electrical equipment for measurement control and laboratory use Following th...

Отзывы: