background image

3.3 GOOD PRACTICE GUIDELINES

 1.

For optimum performance a calibration routine should be carried out at the beginning and end of
every sample batch.

2.

To ensure accurate results are obtained the sample area lid should be kept in the closed position
during measurement.

3.

The styrene cuvettes supplied with the unit are disposable (i.e; ideally they should be used once and
then thrown away). Some repeat use is possible, providing extreme care is taken during cleaning, to
ensure no damage occurs to the polished surface.

4.

Plastic cuvettes are not suitable for use with organic solvents.

5.

Glassware used in the preparation of standards should be made of a high grade borosilicate glass.
The use of soda glass should be avoided wherever possible as leaching can occur during prolonged
contact, giving erroneous results.

6.

Glass cuvettes should be thoroughly cleaned after use. Discard when scratches become evident in
polished surfaces.

7.

Chemical reagents should, wherever possible, be of high grade quality. Contamination can cause
problems, even at very low levels. Diluents (i.e; water or solvents) must be free from impurities.

8.

There are some substances which do not follow Beer’s Law. When attempting a new method it is
advised that linearity checks should be performed over the range of concentrations being used. This
can be carried out by preparing a quantity of known strength solutions and checking the results.

a)

Deviations from Beer’s Law may occur at high concentrations by association of molecular
ionic species.

b)

Deviations from Beer’s Law may occur at low concentrations by variation in hydration,
introducing changes in the nature of complex ions.

c)

Absorption which does not obey Beer’s Law will require a graph of known standards to be
plotted. This should indicate Reading vs Concentration. The reading obtained from theunknowns can
then be related to the concentrations from the graph.

 9.

Samples and standards can “outgas” when left in the cuvette. Bubbles formed on the cuvette

walls will cause reading errors.

6300/REV B/12-96

9

Содержание 6300

Страница 1: ...MODEL 6300 VISIBLE RANGE SPECTROPHOTOMETER OPERATING MANUAL 630 026...

Страница 2: ...tem of work in addition to the detailed instructions specified in this manual 3 The covers on the unit should only be removed by personnel who have been trained to avoid the risk of shock 4 References...

Страница 3: ...N 2 INSTALLATION Unpacking 2 1 Installation 2 2 Displays 2 3 Controls 2 4 Outputs 2 5 SECTION 3 OPERATION Initial Set Up 3 1 Sample Measurement 3 2 Good Practice Guidelines 3 3 SECTION 4 MAINTENANCE G...

Страница 4: ...l contamination Combined with a mechanically rigid structure the 6300 provides a system with fast warm up low drift and high reliability 1 2 INSTRUMENT SPECIFICATIONS Wavelength Range 320 950nm Resolu...

Страница 5: ...your local Distributor before proceeding further Fuse Rating 2A F fast blow type NOTE The unit should be positioned within 1 5 metres of an earthed mains supply VOLTAGE SELECT NOTE When changing the...

Страница 6: ...e Concentration 2 Primary display adjust annunciator 3 Secondary display area Wavelength Factor 4 Primary display units 5 Secondary display adjust annunciator 6 Operation with PC 7 Menu options T ABS...

Страница 7: ...e displayed menu option 4 initiates a calibration routine 5 Print key Provides a printout of the current reading with an incremental sample number When pressed for the first time after a calibration t...

Страница 8: ...mp when replacement is necessary NOTE The Lamp Access Panel and all ventilation slots must not be covered or obstructed at any time 2 ROCKER SWITCH On Off switch for the unit 3 POWER IN SOCKET IEC typ...

Страница 9: ...ot be found the most likely cause is that the sample chamber lid is not closed 3 2 SAMPLE MEASUREMENT Many methodologies are well documented and where these are available the recommended wavelength sh...

Страница 10: ...ctivated to perform the zero setting and this part of the routine is therefore independent of the solution in the light path The following error codes are possible after a calibration Err 1 Dark cal e...

Страница 11: ...ion to the set value by calculating a multiplication factor for the current absorbance If the factor calculated is out of range then the instrument will display Err 3 for 2 seconds after the calibrati...

Страница 12: ...ed surfaces 7 Chemical reagents should wherever possible be of high grade quality Contamination can cause problems even at very low levels Diluents i e water or solvents must be free from impurities 8...

Страница 13: ...urce if this fails Failure should be suspected if lamp failure indicator appears on the display This can be confirmed by looking in the sample chamber The Tungsten Halogen lamp is available from the M...

Страница 14: ...CESSORIES 5 1 OPTIONAL ACCESSORIES The following list of items are available as optional accessories for use with the Model 6300 630 005 Adjustable Cell Holder 630 020 Test Tube Holder 12mm 630 028 Du...

Страница 15: ...ASCII TAB character regardless of the 6300 operating mode For example 0 001 540 ASCII C CR Outputs concentration and wavelength separated by an ASCII TAB character regardless of the 6300 operating mod...

Страница 16: ...active GND 7 Suggested interconnections are detailed below 6300 IBM PC XT 25 way D TXD 2 2 TXD From PC RXD 3 3 RXD To PC RTS 4 4 RTS From PC CTS 5 5 CTS To PC DSR 6 6 DSR To PC DCD 8 8 DCD To PC DTR 2...

Страница 17: ...992 Electromagnetic compatibility Generic immunity standard Performance criterion B EN 61010 1 1993 Safety requirements for electrical equipment for measurement control and laboratory use Following th...

Отзывы: