RESISTANCE MEASUREMENTS 

WARNING:

 To avoid electric shock, disconnect power to the unit under test and discharge 

all capacitors before taking any resistance measurements. Remove the batteries and 
unplug the line cords. 

1. 

Set the function switch to the green         

Ω



CAP 

position. 

Insert the black test lead banana plug into the negative 

COM

 jack. 

Insert the red test lead banana plug into the positive 

 jack. 

2. 

Press the 

MODE

 button to indicate “

"

on the display. 

3. 

Touch the test probe tips across the circuit or part under test. It is best to            

disconnect one side of the part under test so the rest of the circuit will not interfere with 
the resistance reading. 

4. 

Read the resistance in the display. 

 
 

 
 
CONTINUITY CHECK 

WARNING: 

To avoid electric shock, never measure continuity on circuits or wires that have 

voltage on them. 

1.  Set the function switch to the green           

Ω

 

CAP

   position. 

2.  Insert the black lead banana plug into the negative 

COM

 jack. 

Insert the red test lead banana plug into the positive 

 jack. 

3.  Press the 

MODE

 button to indicate”   

"

and “

Ω”

 on the display 

4.  Touch the test probe tips to the circuit or wire you wish to check.  
5.  If the resistance is less than approximately 50

, the audible signal will sound. If the 

circuit is open, the display will indicate “

OL

”. 

 

 

 

DIODE TEST 

1. Set the function switch to the green          

Ω

 

CAP

 position. 

2. Insert the black test lead banana plug into the negative 

COM

 jack and the red    test lead 

banana plug into the positive 

V

 jack. 

3. Press the 

MODE

 button to indicate      and 

V

 

on the



 display. 

4. Touch the test probes to the diode under test. Forward voltage will typically indicate 0.400 

to 0.700V. Reverse voltage will indicate “

OL

”. Shorted devices will indicate near 0V and 

an open device will indicate “

OL

” in both polarities. 

Содержание DT-985

Страница 1: ...DT 985 TRMS MULTIMETER ...

Страница 2: ... ranges that may in normal use be subjected to particularly hazardous voltages For maximum safety the meter and its test leads should not be handled when these terminals are energized This symbol indicates that a device is protected throughout by double insulation or reinforced insulation PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equipment of OVERVOLTAGE CATEGORY I is eq...

Страница 3: ... EXTREME CAUTION when working with high voltages 3 DO NOT measure voltage if the voltage on the COM input jack exceeds 1000V above earth ground 4 NEVER connect the meter leads across a voltage source while the function switch is in the current resistance or diode mode Doing so can damage the meter 5 ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistanc...

Страница 4: ... MIN button 3 RANGE button 4 Mode button 5 Function switch 6 mA µA and 10A input jacks 7 COM input jack 8 Positive input jack 9 HOLD and Backlight button 10 Hz and button 11 RELATIVE button 12 Battery Cover Note Tilt stand and battery compartment are on rear of unit ...

Страница 5: ...cro 10 6 amps cap m milli 10 3 volts amps A Amps k kilo 103 ohms F Farads capacitance M mega 106 ohms Ohms Hz Hertz frequency V Volts Percent duty ratio REL Relative AC Alternating current AUTO Autoranging DC Direct current HOLD Display hold ºF Degrees Fahrenheit ºC Degrees Centigrade MAX Maximum MIN Minimum ...

Страница 6: ... AC VOLTAGE FREQUENCY DUTY CYCLE MEASUREMENTS WARNING Risk of Electrocution The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets As a result the reading may show 0 volts when the outlet actually has voltage on it Make sure the probe tips are touching the metal contacts inside the outlet before ...

Страница 7: ... FREQUENCY DUTY CYCLE MEASUREMENTS CAUTION Do not make 10A current measurements for longer than 30 seconds Exceeding 30 seconds may cause damage to the meter and or the test leads 1 Insert the black test lead banana plug into the negative COM jack 2 For current measurements up to 6000µA AC set the function switch to the yellow µA position and insert the red test lead banana plug into the µA mA jac...

Страница 8: ... circuits or wires that have voltage on them 1 Set the function switch to the green Ω CAP position 2 Insert the black lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive jack 3 Press the MODE button to indicate and Ω on the display 4 Touch the test probe tips to the circuit or wire you wish to check 5 If the resistance is less than approximately 50 th...

Страница 9: ... the batteries and unplug the line cords 1 Set the rotary function switch to the green Ω CAP position 2 Insert the black test lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive V jack 3 Press the MODE button to indicate nF on the display 4 Touch the test leads to the capacitor to be tested 5 The test may take up to 3 minutes or more for large capacit...

Страница 10: ...he meter will display and hold the minimum reading and will update only when a new min occurs 3 To exit MAX MIN mode press and hold the MAX MIN key for 2 seconds RELATIVE MODE The relative measurement feature allows you to make measurements relative to a stored reference value A reference voltage current etc can be stored and measurements made in comparison to that value The displayed value is the...

Страница 11: ...TALLATION WARNING To avoid electric shock disconnect the test leads from any source of voltage before removing the battery cover 1 Turn power off and disconnect the test leads from the meter 2 Open the rear battery cover by removing the screw using a Phillips head screwdriver 3 Insert the battery into battery holder observing the correct polarity 4 Put the battery cover back in place Secure with t...

Страница 12: ...rom 5 of range to 100 of range DC Current 600 A 0 1 A 1 0 reading 3 digits 6000 A 1 A 60mA 0 01mA 600mA 0 1mA 6A 0 001A 1 5 reading 3 digits 10A 0 01A 10A 30 sec max with reduced accuracy AC Current 45Hz to 60Hz 600 A 0 1 A 1 5 reading 3 digits 6000 A 1 A 60mA 0 01mA 600mA 0 1mA 6A 0 001A 2 0 reading 5 digits 10A 0 01A 10A 30 sec max with reduced accuracy All AC voltage ranges are specified from 5...

Страница 13: ... 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999kHz 0 001kHz 99 99kHz 0 01kHz 999 9kHz 0 1kHz 9 999MHz 0 001MHz 10MHz 0 01MHz Frequency electrical 10 00 10KHz 0 01Hz 1 2 reading Sensitivity 15Vrms Duty Cycle 0 1 to 99 9 0 1 1 2 reading 2 digits Pulse width 100µs 100ms Frequency 5Hz to 10kHz Temp type K 4 to 1400 F 1 F 3 0 reading 5 C 9 F digits probe accuracy not included 20 to 760 C 1 C Note Accuracy specificati...

Страница 14: ... times per second nominal Low Battery Indication is displayed if battery voltage drops below operating voltage Battery 1 5VAAA 4 battery Fuses mA µA ranges 0 8A 1000V ceramic fast blow A range 10A 1000V ceramic fast blow Operating Temperature 41 F to 104 F 5 C to 40 C Storage Temperature 4 F to 140 F 20 C to 60 C Operating Humidity Max 80 up to 87 F 31 C decreasing linearly to 50 at 104 F 40 C Sto...

Отзывы: