Installation
19
After these operations, the thermographic image of the selected object
becomes visible on the display. For this, the system uses factory-set
default parameters. (See also description of VARIOSCAN control
software.)
Test and check list
The instructions for the first use of the VARIOSCAN
high resolution
thermographic system are based on basic settings that are necessary
to obtain a first thermographic image.
The test object used was a person. For objects of higher or lower
temperature, you must change the temperature level.
Please note that only a few basic settings are necessary to make a
thermographic image visible. For thermal measurements however,
additional items must be considered (cf. Section 9 -Temperature
measurement).
Before starting operation in normal mode, make sure the following
requirements are met:
1. Dewar flask is cooled down (only on Model 3021).
2. Cover of entrance window has been removed.
3. Power supply unit is connected or a charged battery inserted.
4. Power switch on camera back set to
I
(3021, 3022) or
II
(3021-ST).
InfraTec
Содержание VARIOSCAN high resolution 3021
Страница 1: ...VARIOSCAN high resolution Models 3021 3021 ST and 3022 Manual JENOPTIK Group I n f r a T e c ...
Страница 10: ...Technical Description 10 I n f r a T e c ...
Страница 12: ...Technical Description 12 I n f r a T e c ...
Страница 14: ...Technical Description 14 Total weight of system in carrying case approx 11 kg I n f r a T e c ...
Страница 24: ...Operation 24 To switch off the activated magnifier again double click any of the zoom keys I n f r a T e c ...
Страница 55: ...Maintenance 55 I n f r a T e c ...