Installation
18
How to fill the Dewar flask with liquid nitrogen.
q
Unscrew the sealing cap on top of the camera head.
q
Use the provided funnel and an auxiliary vessel to fill the Dewar
flask slowly with nitrogen.
Note:
q
First fill in only a little amount of nitrogen to avoid boiling over. Fill
the flask completely only, after it has cooled down sufficiently.
q
Screw the cap onto the filling hole. With a fully charged Dewar flask,
you can work for about 10 hours.
Operating the camera
q
Connect power supply unit to camera head or push charged video
battery pack into the battery compartment of the camera until the
lock snaps in.
q
Turn power switch on camera back to
I
(VARIOSCAN 3021, 3022)
or
II
(VARIOSCAN 3021-ST), the indicator LED on the display
lights. During hardware initialisation the display remains dark.
This process takes about 15 s. Subsequently, the software is
initialised. The display shows a corresponding message.
q
Switch position
I
on VARIOSCAN 3021 - ST means on standby for
the internal Stirling cooler. The integrated PC is not running in this
mode.
q
If you want to work in the file system without online camera mode,
press the button at the bottom left.
q
After initialisation, the camera is ready for taking thermograms.
q
Direct the camera to a test object, for instance, a person.
q
Wait until detector has cooled down (on Model 3021-ST up to five
minutes).
InfraTec
Содержание VARIOSCAN high resolution 3021
Страница 1: ...VARIOSCAN high resolution Models 3021 3021 ST and 3022 Manual JENOPTIK Group I n f r a T e c ...
Страница 10: ...Technical Description 10 I n f r a T e c ...
Страница 12: ...Technical Description 12 I n f r a T e c ...
Страница 14: ...Technical Description 14 Total weight of system in carrying case approx 11 kg I n f r a T e c ...
Страница 24: ...Operation 24 To switch off the activated magnifier again double click any of the zoom keys I n f r a T e c ...
Страница 55: ...Maintenance 55 I n f r a T e c ...