Parameter setup and measuring operation
44
LDM52 OEM, Release 003, Revision 000
6.8.3
TRO – Trigger output
The parametrization of trigger output will be set with command TRO.
x
edge
parameterized the edge of trigger signal
0
rising edge (from LOW to HIGH)
1
falling edge (from HIGH to LOW)
2
every edge
y
delay
parameterized the time (delay) up to the measurement in milliseconds msec
Query:
TRO
Set:
TRO x y
Value range parameter x:
0, 1, 2
Value range parameter y:
0 to max 60 000 msec (equal to 1 minute)
active: from 1 msec upward
disabled: 0 msec
Standard:
0 0
Output: Trigger (output) [TRO]: 0, 0
Voltage levels for trigger signals OUT (load resistance 30 kOhm):
Low-Pegel
0 V
High-Pegel
> 2.9 V
Hysterese
0.1 V