Parameter setup and measuring operation
LDM52 OEM, Release 003, Revision 000
25
6.4.10
PR – Parameter setting
Resetting of all parameters to factory settings (default values)
The following parameters are not reset by entering PR:
BR
Baud rate
RS
Serial port
SB
Stop bits
Setting parameters for serial interface
Input:
PR
Output:
Parameters set to firmware defaults.
Baudrate of serial port [BR]:
115200
Stopbits of serial port [SB]:
1
Average [SA]:
1
Measurement frequency [MF]:
0.0
Minimum distance from target [MW]:
-5000000
Maximum distance from target [MW]:
5000000
Offset in 'mm / 10' [OF]:
0
Parametrization of switching output Q1 [Q1]:
0, 1000000, 2500, 0
Parametrization of switching output Q2 [Q2]:
0, 1000000, 2500, 0
Parametrization of switching output Q3 [Q3]:
0, 1000000, 2500, 0
Parametrization of the analog switching output QA [QA]: 0, 1000000
Unit for the distances [MUN]:
mm
Trigger (input) [TRI]:
0, 0
Trigger (output) [TRO]:
0, 0
Autostart commands [AS]:
DT
Output format [SD]:
0 0 0 0
Terminator [TE]:
0x0D0A
Scale factor [SF]:
0.000
Error mode [SE]:
0
Separator [SP]:
0x2C
Heating on threshold [HE]:
4
Heating off threshold [HE]:
10
Standard tracking mode from menu [MCT]: 0