22
If the relay action is set to
LO
W and the hysteresis mode is EDGE,
the relay will turn
ON
at [(RELAY SET POINT)], and will turn
OFF
at
[RELAY SET POINT+ (hysteresis value)].
CAUTION :
Setting any hysteresis value to zero may cause jitter and possibly
damage the relay(s).
D. pH RELAYS
There
are
two
independent
relay
channels for pH display
which has
independen
t
set point and control
action
(HIGH or LOW,
see
figure
1.).
The
hysteresis mode (center
or edge, see figure 1
.
)
and hysteresis value will
be used by
both
pH
relays.
The action of the pH
relays is dependent on
set point, relay control
action (HIGH or LOW),
hysteresis mode (Center
or Edge), hysteresis
Value and the current
pH display (See figure 1.)
Figure 1
hysteresis
hysteresis
LOW
SET
POINT
HIGH
SET
POINT
POINT
SET
HIGH
POINT
SET
LOW
hysteresis
hysteresis
EDGE
EDGE
EDGE CONTROL
ON
ON
OFF
OFF
ON
ON
ON
OFF
ON
OFF
OFF
OFF
OFF
ON
ON
OFF
CENTER CONTROL
High set ON/OFF control
Low set ON/OFF control
Low set ON/OFF control
High set ON/OFF control
pH/ORP
pH/ORP
pH/ORP
pH/ORP