24
VII. SPECIFICATIONS
TURBIDITY
Probe
Mode
Range
NTU(FNU)
Resolution
NTU(FNU)
Accuracy
NTU(FNU)
Auto
Washing
Time(min)
TC-100
0.00 ~ 100.0 0.01
±3% F.S.
30
TC-500
0.0 ~ 500.0
0.1
±3% F.S.
10
TC-3000
0 ~ 3000
1
±3% F.S.
10
SUSPENDED SOLIDS
Probe
Mode
Range
mg/L
Resolution
mg/L
Accuracy
mg/L
Auto
Washing
Time(min)
TCS-1000 0.0 ~ 1000.0 0.1
±3% F.S.
10
TS-MXS-A 0 ~ 50000
1
±3% F.S.
10
4-20 mA Analog Output
Current output range
4 to 20 mA (isolated)
Current output scale
user programmable
Maximum load
500 Ω
Isolation voltage
500 VDC
Controller
Control type
(two) ON/OFF control
Relay output
5 A at 115 VAC or 2.5 A at 220 VAC
Resistive load only
External Cleaning Controller
Control type
ON/OFF control
ON/OFF Time
ON 0 ~ 9999 sec or OFF 0 ~ 9999 sec
GENERAL
Keys
Audio feedback in all keys
Security protect
4-digit password
Communication
RS485,
with MODBUS
communication protocol.
Power:
115 VAC or 230 VAC 50/60 Hz
Fuse
0.315Amp/250V fast acting glass tube
Ambient Temperature range
0.0 to 50.0 °C
Display:
128 x 64 graphic LCD w/ backlight
Case
IPT65 ¼ DIN case, depth 148 mm
Weight
950 g