VI. 4 - 20 mA ANALOG OUTPUT
A. ISOLATION VOLTAGE
The maximum isolation voltage of the 4-20 mA output is 500 VDC.
The voltage differential between the 4-20 mA output and the load
should not exceed 500 VDC.
B. OUTPUT LOAD
The maximum load is 500 . Output current inaccuracies may occur
for load impedance in excess 500 .
C. TURBIDITY AND SUSPENDED SOLIDS LINEAR OUTPUT
The analog output will produce a linear analog output if the user
selects this option . The analog output will be dependent on the
Turbidity or Suspended Solids_4 mA setting, Turbidity or
Suspended Solids_20 mA setting and the current Turbidity or
Suspended Solids display.