27
Ergo-Flex
Operating / Service Manual
TROUBLESHOOTING
Problem / Condition
Solution
Possible Cause
Fiberoptic cables will not fit onto
scope.
Screw-on appropriate adaptor.
See pages
6-7.
Incorrect adaptor on light post on scope.
Focus ring on scope is not adjusted
properly.
Set focus ring to align with dot for
no optical correction. Rotate focus
ring to achieve best focus. Varies
per user. See page 8.
Image observed through scope is
out of focus (blurry).
Debris on tip of scope.
Fiberoptic cable not fully inserted into
light source or scope connection.
Broken fibers in fiberoptic cable.
Debris or fingerprints on eyepiece.
Clean tip of scope.
Verify good connections.
Replace fiberoptic cable.
Clean eyepiece.
Image observed through scope is
dim.
Debris on scope tip covering light
emitting window(s).
Improper lamp in light source.
Improper heat filtering of fiberoptic light
source.
Clean scope tip.
Install proper lamp.
Check for broken or missing heat
filter in fiberoptic light source.
Scope tip becomes warm.
Internal angulation mechanism is out-
of-adjustment.
Return to JEDMED for service.
Angulation of tip is limited, not
symmetrical, requires effort, or does
not return to center when released.