DEVICE OPERATION - Ventilation Cap
1. Check venting connector - ensure connector had no scratches, defects or residues
2. Check leak tester - check for cracks, scratches, defects or residues.
- squeeze hand pump bulb (verify air exits connector)
Check Components
Tester connection cap
Air should exit here
Hand pump
DEVICE OPERATION - Leak Testing
To avoid damage to scope, perform leak test
before every cleaning, disinfection, sterilization or
immersion in any liquid.
CAUTION
A ventilation cap is used to open an airway to the inside of
the scope. When the cap is connected to the scope, internal
and external atmospheric pressures are allowed to equalize.
Pressure equalization is required when transporting the scope
via air-cargo or if placing scope in any low-pressure chamber.
Without venting, internal scope pressure could cause insertion
tube sheath to expand and rupture.
Installing ventilation cap:
- Align cap to connector on scope.
- Gently push inward while turning 1/8 turn.
Reverse procedure to remove cap.
Ventilation cap will allow fluids to enter scope.
Ventilation cap MUST be removed from scope
before immersing scope in any liquid.
CAUTION
Scope is vented when
ventilation cap is installed.