Jedmed EF-N Скачать руководство пользователя страница 10

Ergo-Flex

Operating / Service Manual

 

10

DEVICE OPERATION - Light Source Risks

Background

High-intensity light sources used for endoscope 

illumination may pose multiple risks to the patient 

and/or user.  Primary risks include patient burns, fire 

and electrical shock.  It is important to understand 

and limit these risks in all cases.

Many types and manufacture of fiberoptic light 

sources are readily available.  Common types include: 

halogen (incandescent), light-emitting diode (LED), 

metal-halide, and xenon-arc.  In general, all light 

sources and cables must be medical-grade to ensure 

patient safety.

Heat / Burn Hazards

Each light source emits different amounts of heat 

associated with the light, depending on the nature 

of the light and type of construction.  LED sources 

typically emit the least heat, while xenon-arc sources 

generally emit the most.  

Emitted light may cause the tip of the scope to heat, 

at times, significantly.  Additionally, the emitted light 

may be absorbed by tissue or material upon which it 

is shining, also causing significant heating.  Broken 

or  missing  heat  filters,  incorrect  lamps  in  sources 

and fiberoptic cable diameter can also affect heat at 

the distal tip.

Before use, be sure to test any light source with the 

scope to determine that the tip temperature remains 

as cool as possible (ideally under 41º

C/106

ºF) and 

that the emitted light does not cause excessive 

heating of tissue.  Tip temperatures exceeding 

41º

C/106

ºF may be allowable, in some instances, 

depending on location of viewing, type of tissue 

contact and duration of use, among others.  

In all cases, it is imperative to assure that tip heat 

and/or emitted light does not cause an unacceptable 

risk to the patient or user.

Due  to  the  extreme  nature  of  arc-lamps,  JEDMED 

does not recommend using xenon-arc sources with 

this endoscope.

Fire Hazards

Along with patient burns, extreme light output at the 

tips of endoscopes can also be a fire hazard.  Always 

keep scope tip away from flammable materials and 

turn off light source when not in use.

Never leave light operating while unattended.

Electrical Shock Hazard

Leakage current is always present on some level 

in mains (ac) powered equipment.  In some cases, 

leakage current can be felt by the patient as a tingling 

sensation on sensitive tissue such as mucus-lined 

nasal passages.  To minimize any possibility of 

macro or micro shock from fiberoptic light sources, all 

sources and fiberoptic cables used must be Type BF 

and medical-grade.  BF indicates that each device is 

“floating” and not electrically connected to the patient 

in any fashion.

Explosion Hazard - Do NOT use in the presence of 

flammable anesthetics!

WARNING

!

Содержание EF-N

Страница 1: ...ervice Manual An Employee Owned Company IMPORTANT Read all instructions before assembling or using this device Instructions for Use Laryngo Nasopharyngoscope 49 4930 EF N 3 4mm 49 4928 EF N Slim 2 8mm...

Страница 2: ...s quality system has placed JEDMED as an industry leader As always our focus remains on you our valued customer Building rewarding and lasting relationships is the cornerstone of JEDMED s business phi...

Страница 3: ...ion 9 Light Source Risks 10 List of Warnings 11 Functional Test Before Using 12 Ventilation Cap Leak Testing 14 Cleaning Disinfection 17 Maintenance Storage Transport 19 Reorder Information 20 Specifi...

Страница 4: ...hen using fiberoptic light sources keep flammables away from light port fiberoptic cable ends and endoscope tip Disconnect from power before cleaning servicing or storing All cleaning and disinfecting...

Страница 5: ...e designed for examining the larynx nasal cavity and nasopharynx These areas encompass the upper respiratory tract from the nasal passages to the vocal chords Ergo Flex scopes are intended to be used...

Страница 6: ...bundle Handle Leakage tester connector Fiberoptic cable connector Control lever to bend the distal tip Focus ring Eyepiece Insertion tube with bendable distal tip Features Package Contents Scope Leak...

Страница 7: ...cable to suitable fiberoptic light source Use only medical grade fiberoptic light sources and light cables to assure patient safety See page 10 11 Warning Karl Storz adaptor removable Wolf adaptor rem...

Страница 8: ...Scope MUST be cleaned and disinfected before initial and every use WARNING Users wearing glasses or contact lenses corrected vision Users who remove glasses or contact lenses uncorrected vision Rotat...

Страница 9: ...endocoupler Next connect the endocoupler to the scope at the common B cup eyepiece interface Rotate focus ring to align its dot to dot on scope body Focus as needed using focus ring on endocoupler Co...

Страница 10: ...ith the scope to determine that the tip temperature remains as cool as possible ideally under 41 C 106 F and that the emitted light does not cause excessive heating of tissue Tip temperatures exceedin...

Страница 11: ...fiberoptic light sources video cameras or other equipment attached to the scope MUST be medical grade and type BF floating or electrically isolated connections In addition ALWAYS use an isolated fibe...

Страница 12: ...tube and distal tip to verify both are smooth and unblemished 5 Test eyepiece focus adjustment 6 Place scope near representative anatomical target and verify colors observed by naked eye under ambient...

Страница 13: ...13 Ergo Flex Operating Service Manual Blank...

Страница 14: ...o the inside of the scope When the cap is connected to the scope internal and external atmospheric pressures are allowed to equalize Pressure equalization is required when transporting the scope via a...

Страница 15: ...s slightly while inflated 3 Observe gauge If pressure reading drops more than 10 mmHg within 1 minute DO NOT USE scope must be returned to JEDMED for service see page 25 If pressure remains constant a...

Страница 16: ...Ergo Flex Operating Service Manual 16 Blank...

Страница 17: ...medical grade forced air until completely dry 4 Functionally Test Scope ENZOL Enzymatic Detergent Do NOT use ultrasonic cleaner STORAGE Step 5 1 Assure Scope is Completely Dry 2 Store Scope in a cove...

Страница 18: ...Blank...

Страница 19: ...ue to reduced atmospheric pressure Refer to page 14 Transport Scope and accessories should be thoroughly cleaned disinfected and dried before storing Keep the instrument lying flat or hanging with the...

Страница 20: ...for replacement parts NOTE Replacement Accessory Items Order Description 49 4980 Leakage Tester 49 4985 Ventilation Cap 49 4172 Karl Storz Adaptor 49 4170 Wolf Adaptor 49 4104 ACMI OLYMPUS Adaptor fo...

Страница 21: ...king Length 320mm Bend Angle up down 130 Bending Radius 8mm Optical Field of View 80 Depth of Field 1 50mm Weight 230g 8 oz EF N Slim Sheath Diameter 2 8mm Working Length 320mm Bend Angle up down 130...

Страница 22: ...Ethicon Inc a Johnson Johnson Company Steris is a registered trademark of STERIS Corporation Product Trade Name High Level Disinfectant Detergent CIDEX OPA CIDEX Activated Dialdehyde Solution 2 4 glu...

Страница 23: ...SERVICE INFORMATION...

Страница 24: ...parts l Before returning to JEDMED for service instrument must be cleaned disinfected and labelled accordingly If scope has failed leak testing and cannot be immersed for cleaning see next page for re...

Страница 25: ...he problem and contact information Return to JEDMED for service A protective plastic sheet is supplied for return of soiled flexible endoscopes 1 Returning Endoscopes for Service An Employee Owned Com...

Страница 26: ...Blank...

Страница 27: ...t fully inserted into light source or scope connection Broken fibers in fiberoptic cable Debris or fingerprints on eyepiece Clean tip of scope Verify good connections Replace fiberoptic cable Clean ey...

Страница 28: ...NOTES...

Страница 29: ...5416 JEDMED Court St Louis MO 63129 314 845 3770 phone 314 845 3771 fax e mail info jedmed com website www JEDMED com...

Отзывы: