![JDS Uniphase OLP-85 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/jds-uniphase/olp-85/olp-85_user-manual_3631986062.webp)
62
OLP-85
9 R
EMOTE
C
ONTROL
C
OMMANDS
Setups for the Probe application/PCM application
Command string
Parameter type / Response type / Unit / Info
:APP:ID:VSBL
<STRING>
Makes the specified application visible (i.e. brings the specified
application to foreground).
Notes:
Full functionality is only available for the active
application, so it may be important to know which
application is active.
Valid values:
All application IDs returned by :APPM:LIST?.
:APP:ID:VSBL?
Returns the name of measurement application which is currently visible
on the display.
Response type:
<STRING>
See :APP:ID:VSBL command for more information.
:AUTO:CNTR
<BOOLEAN>
Controls fiber center tracking for live video images in high-
magnification mode.
Default:
TRUE
:AUTO:CNTR?
Returns TRUE if the fiber center tracking in live video is turned on.
Response type:
<BOOLEAN>
See :AUTO:CNTR command for more information.
:CAPT:BTTN:MODE
<NR1>
Controls the function of the probes capture button.
Valid values:
0: Freeze
1: Test
Default:
1
:CAPT:BTTN:MODE?
Returns which function the probes capture button has.
Response type:
<NR1>
See :CAPT:BTTN:MODE command for more information.
:FQM:ENAB
<BOOLEAN>
The Digital Probe calculates a focus quality metric for each video frame.
If this flag is set to TRUE, the result is displayed as a focus bar on the
screen. Deactivating this calculation can speed up video frame rate in
some cases.
Default:
TRUE
Содержание OLP-85
Страница 1: ......
Страница 88: ...88 OLP 85 ROHS...