ConsoleP
edal
Times
Pre-Set
P1 Low
Pre-Set
P2 Medium
Pre-Set
P3 High
t1
30 sg
40 sg
60sg
t2
15 sg
25 sg
40sg
t3
4 min
4 min
4 min
4
MAIN MENU
PRE-SET PROFILES
Basic Mode
The unit will keep the selected temperature at the point where the Control TC is placed. The maximum
working time and the maximum ramp-up rate are set in the station settings menu. In the station settings
you can also select the maximum temperature available and set a unique temperature.
Power Mode
The unit works with the power selected. If you wish to see the actual temperature, plug in the Control TC.
Builder Profile
Build your custom profile step by step, test it and check it as many times as you wish. Once defined, save
it in the profile memory.
Pre-Set Profiles
Run a pre-set profile. The profiles 1 (low), 2 (medium) and 3 (high) are pre-set for a standard rework process.
Custom Profiles
Run a saved custom profile. The custom profiles must be previously set by means of the builder profile or
by using the options of the Profile Settings Menu: Edit, Copy and Delete.
The unit incorporates 3 default profiles which are appropriate for most common jobs on FR4 bilayer
substrates, multilayer and thick multilayer respectively.
Profiles are appropriate if you work with
PH321
support at its lowest working position (21mm *) and with
the PCB focused on the PH. The Control TC must be placed with Kapton on the top side of the board,
near the component to be repaired.
Control TC
21mm*
PCB
PH321 Support
PH
T1sg
T2sg
P1
P2
P3
T3sg
100ºC
80ºC
Tamb