To ensure safe operation and long term performance stability, it is essential that you follow
this maintenance instruction all the time.
The performance test should be done every 2 years to ensure that the device operates
properly and safely at all times. It can be performed either by the manufacturer or by an
authorized dealer according to the regulations governing manufacturing of medical products.
1. REPLACING THE CUFF COVER
Remove the old cuff cover as follow;
Remove the rubber molding which is located in the front of the arm insertion part.
Press the start button, then the cuff will be wrapped.
When the cuff is wrapped approximately half, turn the power off.
The cuff cover will be strained.
Detach the cuff cover from the magic tape.
(The location of the magic tape is as the picture.)
Pull the upper part of the cover down direction, and
then pull out the ring of it very gently.
Turn the power on again, then the cuff will
return the original status.
Pull out the rear ring.
Remove the old cuff cover completely.
Replace the new cuff cover as follow;
Place the new one through the arm insertion part. At this time, the large ring is for front,
and the small ring for rear. (The sewed part of the cover should be placed bottom.)
Press the start button and turn the power off again for half wrapping of the cuff.
Push the cover’s rings into the body.
Attach the cuff to the magic tape with pressing.
Fix the rubber molding like the beginning.
Note
If the cuff cover is not placed properly, it may cause problems in operation.