2. MEASUREMENT

Check the voltage and turn the power on

With the switch-on, a music sound flows as the

LED screen is activated

But, the SOUND ON/OFF switch in the rear of the

unit should be set as  ON

For the card users, proceed with checking after

entering the card into the card-reader.

When the card is inserted, a voice message

comes out to “press the start button”

Put your right or left arm into the cuff
deeply. For FT-500R, measure with right
arm, for FT-500L, measure with left arm.

Note

the CARD ON/OFF switch in the rear of the unit should be set as  ON
if the card is damaged or expired, the voice message comes out as “Can not use ID
card.” In this case, purchase new one and try again.

Note

An optimal arm circumferance for this equipment is 9 to 14

FT-500L

FT-500R

Содержание FT-500L

Страница 1: ......

Страница 2: ...ER 9 3 REAR PART 10 4 ACCESSORIES 12 5 OPTIONS 12 1 CONNECTING POWER CABLE AC INPUT 13 2 ASSEMBLING OF ARM REST 13 3 LOADING THE PRINT PAPER 14 4 CONNECTING PORTS RS 232C 14 5 SETTING TIME AND DATE 15 1 CAUTION FOR MEASUREMENT 16 2 MEASUREMENT 17 3 DATA ON MEASUREMENT BY PRINTER 20 21 1 AFTER SERVICE 24 2 PACKING AND TRANSPORT 24 25 26 ...

Страница 3: ...e monitor armed with the belt type cuff that facilitates the test time quicker and provides the results more reliable 2 WORD DEFINITIONS To ensure safe operation and long term performance stability it is essential that you fully understand the functions operating and maintenance instructions by reading this manual before operating your unit Particular attention must be paid to all warnings caution...

Страница 4: ...utor 2 FT series are specified as Class 1 type BF unit under the standard of IEC 60601 1 Safety of Medical Electrical Equipment Therefore patients must not touch or handle inner side of the system at any time 3 Do not modify the unit If any modification is desired ask JAWON or its authorized dealer for service 4 The unit has been factory adjusted for optimum performance Do not attempt to adjust an...

Страница 5: ...failure of user to maintain the unit relieves the manufacturer or his agent of the system s non compliance with specifications or responsibility for any damage or injury This manual is made for informational purpose and this manual and product are not meant to be a substitute for the advice provided by your own physcian or other medical problem You should not use this information contained product...

Страница 6: ...spectively Identifies equipotential ground located next to ground stud on rear panel IEC 417 5021 Identifies the point where the safety ground of the system is fastened to the chassis Do not open This is for factory only CAL 5 SAFETY SYMBOLS The International Electrotechnical Commission IEC has established a set of symbols for medical electrical equipment which classifies a connection or warnning ...

Страница 7: ...ed due to high pressurizing or irregular operation is done press this button then the cuff will be exhausted rapidly PRINTER COVER It protects the printer AUTOMATIC CUTTER Printer paper let out slot Printing paper is automatically cut off to let it out through the slot CUFF It wraps and releases the arm automatically for measurement ARM REST When the arm is placed on the cuff the arm supporter sus...

Страница 8: ...ttached to FT 750AF R L model ID card can be issued either by the machine manager or by the manufacturer of the model The card can store ID number and measured results The card stores two previous measured results and can contain three measured results with the current one altogether When the model manager writes the card please refer to the manual and specifications for operation and programs att...

Страница 9: ...f all memorized data would be deleted When you set the date and time use as a key which increases the number FEED button Use this button for setting the paper When you set the date and time use as a key which decreases the number SET button set the date and time The functions are as follows when press this before or after measurement It does n t work during measurement Sequence is HOUR MIN MON DAY...

Страница 10: ...een the main body and a computer or other equipment with cable RS 232C to transfer the data collected or measured second for connecting between the main body and the coin slot with RS 232C calble to transfer the data SENSOR ON OFF Human sensor is turned on and off with the switch SOUND ON OFF Music and voice output functions are activated with the switch placed to ON and vice versa CARD ON OFF ID ...

Страница 11: ...INFORMATION BOARD FIXER VOLUME ANNOUNCE ON OFF CARD ON OFF SOUND ON OFF SENSOR ON OFF PRINT ON OFF REAR PART EARTH POTENTIAL EQUALIZATION TERMINAL CAL AC INPUT POWER FUSE PORT RS 232C ...

Страница 12: ...4 ACCESSORIES Manual Poster Guide Arm rest 5 OPTIONS Cart Chair Card reader Card writing device Cards Printer paper Printer Human sensor Earth Power Fuse ...

Страница 13: ...on the rear and turn the POWER ON OFF switch O I on the lower part of the rear See the picture 2 ASSEMBLING OF ARM REST The arm rest is divided into two parts the first one is built in the unit as it is packed in factory For assembling fit the other to the one as below See the picture ...

Страница 14: ...2C To transmit the data with connecting a computer or other external options to the unit Connect the RS 232C cable both to port of the unit and to the computer jack or other external options See the picture Note Being thermal type printing is photocopied on one side of the paper slippery side without using printing ink Please check remainder of the paper always and then replace it Please use exclu...

Страница 15: ... says its counts that have been measured since keeping button ON Press SET button one more First 2 figures will blink To set the current hour press PRINT button to make the number goes up or press FEED button to make the number goes down HOUR After setting the hour press SET button again In this time last 2 figures will blink As the same way as above set the current minute with PRINT and FEED butt...

Страница 16: ...r legs crossed When you fold your sleeves up please make sure not to press your arm Put your arm deeply in order to fit to the arm rest If it is not easy to check your pulse with the stethoscope the measurement cannot go on If you strain to your arm there will be some differences Please relax your arm and have a seat correctly Before measurement take it easy and have a rest for a while ...

Страница 17: ...der When the card is inserted a voice message comes out to press the start button Put your right or left arm into the cuff deeply For FT 500R measure with right arm for FT 500L measure with left arm Note the CARD ON OFF switch in the rear of the unit should be set as ON if the card is damaged or expired the voice message comes out as Can not use ID card In this case purchase new one and try again ...

Страница 18: ...gher than actual value and vice versa Note When the measurement is started the voice message is announced as Starting measurement don t move or speak please When you feel painful and want to stop the measurement press STOP button When the measurement completed the cuff is automatically deflated and it returns to normal condition Simultanously the voice message is announced as Measurement completed...

Страница 19: ...RE is SYSTOLIC 000 DIASTOLIC 000 PULSE 000 Note When the PRINT ON OFF switch on the rear is set as OFF the result will not be printed even if the measurement is completed Pull your arm out Note For the card users two previous results stored in the card can be recalled to compare with the current ones newly checked Three results altogether could be printed out ...

Страница 20: ...r leaks message appears on the printer as ERROR CUFF When the message is repeated call for maintenance service When a testee moves or speaks while in testing message appears on the printer as ERROR MEASURE Try to re test after for a while If the message is repeated call for maintenance service A buzzer goes off once when error takes place then the machine is ready for testing When the machine does...

Страница 21: ...power off The cuff cover will be strained Detach the cuff cover from the magic tape The location of the magic tape is as the picture Pull the upper part of the cover down direction and then pull out the ring of it very gently Turn the power on again then the cuff will return the original status Pull out the rear ring Remove the old cuff cover completely Replace the new cuff cover as follow Place t...

Страница 22: ... do not mount it close to the gas pipe water pipe lightening rod and connection of telephone When you take out the plug grasp the plug exactly Keep the unit dry and do not wet it Never immerse the main unit in the water It could cause damage the electronic parts inside Avoid direct sunshine humidity dust and extreme changes in temperature Remind the permitted current 50VA ...

Страница 23: ...very 2 years Do not put heavy things on the unit Check the unit conditions occasionally Do not use the unit under the condition of shock or vibration Do not install and shock with chemicals or gas When you re use after a long time please check the unit carefully When some problems are found please call our service center Wipe the unit with a damp cloth Do not use benzine alcohol or liquid similar ...

Страница 24: ...ll assist your appropriately qualified technical personnel to repair those parts of unit which are designated by JAWON as repairable How to contact JAWON MEDICAL CO LTD Write us at JAWON MEDICAL CO LTD SEOUL OFFICE JEONGJU BLDG 7F 1451 38 SEOCHO DONG SEOCHO KU SEOUL KOREA TEL 82 2 587 4056 FAX 82 2 588 1937 FACTORY 13 LOT 21 BLOCK JINRYANG INDUSTRIAL COMPLEX KYUNGSAN CITY KYUNGSANG BUKDO 712 830 K...

Страница 25: ...EC AC120V 60Hz US Power consumption 50VA Ambience for operation Temperature 10 40 Humidity Less than 85 non condensing Ambience for storage Temperature 5 50 Humidity Less than 85 non condensing Data transmission RS 232C Dimensions 480 W 310 D 290 H mm Weight Approx 11kg Measuring parts FT 500L with left arm FT 500R with right arm Note 1 It is convenient and useful if you purchase a cart and a chai...

Страница 26: ... NO Note When you receive this warranty make sure that the name of the dealer and the month day and year of purchase are all completed This warranty will not be reissued please keep it in a safe place Date Defection Confirmation Date of purchase Month Day Year Customer Name TEL Address Dealer Name TEL Address ...

Страница 27: ......

Страница 28: ...www jawon co kr ...

Отзывы: