44
8.2 “10 ~ 50mL/min” tubing arrangement and solvent replacement
method for reference cell
Connect the tubing from the mobile phase solvent bottle to port No3. Use
tubing with an internal diameter of 0.8mm and an outer diameter of 1.6mm.
Connect the syringe to port No.6. To replace solvent in the reference cell, draw
up mobile phase solvent by pulling the syringe plunger. Make sure the syringe
is connected during analysis. Connect the tubing from the column to port No.5,
and connect the tubing for waste to port No.2. The refractive index output
signal is reversed. Change the Polarity setting. (Refer to section 4.1.3)
Figure 8.2 “10 ~ 50mL/min” tubing arrangement
Tubing OD=1.6mm, ID=0.8mm
Replace solvent in the reference cell
by sucking the mobile phase solvent.
Keep the syringe connected during
analysis in order to maintain solvent in
the reference cell and to facilitate
solvent replacement.
Tubing
OD=1.6mm
ID=0.8mm
(Same tubing as used to
connect ports No.1-2 or
ports No.4-5. for the
“
~
10mL/min” tubing
arrangement.
Tubing
OD=1.6mm
ID=0.8mm
Sample
Sample IN
Reference
IN
Reference
Out
1
Sample Out
Sample IN
Mobile phase
2
3
4
5
6
1
2
3
4
5
6