INSPECTION AND TROUBLESHOOTING

INSPECTION OF YOUR OPTICELL SOFT PLUS AIR MATTRESS SYSTEM

An inspection plan should be established

 in order for your air mattress to function and last 

for as long as possible. Please contact us if you would like assistance with this. Järven Health 
Care’s contact details can be found on the back cover of this guide.

INSPECTION OF THE MATTRESS AND PUMP

•  Check the electrical cord and plug for signs of damage or abnormal wear.

•  Check the mattress surface for signs of damage or abnormal wear.

•  Check that the cells of the mattress are correctly connected to the air hoses.

•  Start the pump and examine the air flow from the pump (with the air hoses 

disconnected). The air flow should alternate between the two outer holes each cycle 
time. The middle hole should release no air.

•  Check the air hoses for signs of damage or abnormal wear.

•  Check the air hose connector (see page 10).

INSPECTION OF THE OPTITEX COVER

Regularly check the cover for visible signs of damage or wear,

 such as tears, cracks, holes, 

stains or discolouration. Check all sides including the underside.

TROUBLESHOOTING

The pump does not work or does not start:

•   Make sure the electrical plug is properly inserted, then turn on the pump’s power switch.

•   If the power indicator is off, there may be a problem with the electrical socket. Test another 

power outlet.

•   If the power indicator is on but the pump does not work, contact Järven Healthcare.

The mattress does not inflate properly (pressure too low):

•  Set the pressure to maximum, check if the air hoses are pinched anywhere or if there is 

air leakage from any cell. Check that the CPR valve is closed.

Poor air flow:

•  A dirty air filter can reduce air flow. Wash the air filter

 

with mild detergent. The air filter is 

easily accessible on the back of the pump and can be released with a simple motion.

Protective cover 

for air filter

CLEANING

CLEANING THE PUMP

To clean the pump, use a damp cloth and mild detergent on the pump housing and the 
power cable.

CLEANING THE OPTITEX COVER

The cover is made of a polyurethane (PU) material on both top and bottom. The entire cover 
(both upper and lower parts) can be cleaned with disinfectants. Use alcohol-based disinfectants.

Healthcare textiles (mattress coverings, bedding, etc.) can be contaminated by pathogenic 
microorganisms. These should be washed at a minimum of 

70° C

 for at least 10 minutes 

with unscented detergent, according to standard 

”BS-EN 14065: 2016 Textiles – Laundry 

Processed Textiles - Biocontamination Control System”

.

The OptiTex cover can be machine washed at a maximum of 

95° C

.

CLEANING THE CELLS AND AIR HOSES

The cells can be washed separately or in groups using a damp cloth and mild detergent.  
The air hoses can also be wiped with a wet cloth and mild detergent.

CLEANING THE AIR FILTER

The pump has an air filter located on the back. The filter can be easily accessed through a 
protective cover. See picture below.

The filter can be washed with mild detergent.

Содержание OptiCell Soft PLUS

Страница 1: ...1 USER GUIDE OptiCell Soft PLUS Constant low air pressure overlay mattress...

Страница 2: ...important to the user Risk of damage to mattress Important information denoted by this symbol must be observed in order not to damage the mattress or other material CE MARKING J rven Health Care hygi...

Страница 3: ...Categories 1 3 low risk to very high risk for patients within the weight range 30 230 kg The product is intended for use with a sheet between the patient s skin and the mattress surface The mattress...

Страница 4: ...01 EN 597 1 and EN 597 2 SYSTEM FEATURES The OptiCell Soft PLUS air mattress is constructed so that individual cells can be disconnected even while a patient is lying on the mattress The mattress has...

Страница 5: ...re We recommend that the mattress be kept rolled up in one of our optional transport bags The mattress should not be treated carelessly when stored as it risks being damaged Avoid exposing the mattres...

Страница 6: ...ny problems with the mattress s air pressure The test ensures that the patient does not bottom out How the hand check is performed The hand check should always be performed when the patient is positio...

Страница 7: ...on but the pump does not work contact J rven Healthcare The mattress does not inflate properly pressure too low Set the pressure to maximum check if the air hoses are pinched anywhere or if there is a...

Страница 8: ...mately 15 minutes When the lamp is lit the mattress is ready to use UNLOCK The panel can be locked to prevent unauthorised people from changing the system settings To unlock the panel press and hold t...

Страница 9: ...slation EU Directive 85 374 EEC Swedish Law SFS 1992 18 establishes the manufacturer s importer s or seller s statutory liability for any damage caused by the product The regulations apply to physical...

Страница 10: ...ent regarding pressure relief and must be conducted by staff with adequate knowledge and education Based on the patient s assessed status the appropriate mattress is chosen for the correct purpose and...

Страница 11: ...0 824 10 Email healthcare jarven se OptiCell Soft PLUS user guide All trademarks are the property of their respective owners J rven AB reserves the right to update this guide if necessary and or on fi...

Отзывы: