background image

9

USE OF PRODUCT

RECEIVING AND UNPACKING

The OptiCell Soft PLUS mattress is delivered rolled up  
in cardboard. Use of knives should be avoided when  
unpacking and handling due to risk of damage.

An OptiCell air mattress system contains:

•  1 OC-15 pump
•  1 power cable
•  1 mattress, with air hose cable, CPR valve and OptiTex cover

Check that the product is not damaged. In the event of transport damage, see page 17.

The mattress can be set in place and made ready for use immediately.

PLEASE NOTE:

•  The air pump should not be used in humid areas nor should it be left outdoors in damp 

weather conditions. If the pump is exposed to water, avoid handling it until it has dried.

•  Take care when handling the pump, power cable and air hoses. Air hoses can be pinched  

and damaged by bed rails etc.

•  The mattress’s comfort and function change depending on its air pressure in relation to the 

patient’s weight. Keep in mind that a tall patient can weigh the same as a short patient, but 
may need a different pressure. This must always be carefully observed for each patient.

•  The biggest risk for the patient is to “bottom out” on the mattress. The basic rule is that the  

part of the body to be unburdened must not be subject to pressure. If the patient “bottoms 
out”, the objective of the mattress is lost. See page 11 for information on the Hand Check test.

•  The patient should not be positioned with their respiratory system turned towards the mattress.

It is important that no sharp objects come in contact with the mattress. 

 

To prevent damage to the mattress avoid knives, keys and other objects with a cutting 
or puncturing ability from coming into contact with the mattress.

STORAGE

The OptiCell Soft PLUS air mattress system should be stored in the dry and at room temperature.

 

We recommend that the mattress be kept rolled up in one of our optional transport bags.  
The mattress should not be treated carelessly when stored, as it risks being damaged. 

Avoid exposing the mattress to direct sunlight for prolonged periods.

8

The mattress can be used immediately after delivery and unpacking. Inflating the mattress 
takes about 15 minutes and should be done before the patient is placed on the mattress.

WORKING MODES

OptiCell Soft PLUS is a static air pressure mattress with two different working modes:

•  “Care” 

(also called Auto-firm mode) provides maximum pressure to facilitate nursing 

care (

The lamp at position D, page 14, is on

).

•  “Static” 

mode is active when ”Care” mode is not selected. (

The lamp at position D, page 

14, is off

).

In static mode, when the patient is lying down, enter the patient’s weight as an initial 
guideline.

The pump display shows the weight to which the air pressure has been set.

The air pressure can then be raised or lowered according to the patient’s wishes.

If the patient is in a sitting position, increase the weight setting on the pump. For users over 
200 kg, set the pump to maximum pressure (250 kg) at the sitting position.

LOW PRESSURE ALARM

If the mattress air pressure falls too low, the 

“        ” 

lamp lights up and an audio alarm starts.

To respond, start by pressing the 

“Alarm” 

button 

(position E on page 14) 

to silence the alarm 

and then check the following points to find the cause.

•  Check that the air hose is connected correctly to the pump.

•  Check that no air is leaking from the valves connecting the air cells and air hose.

•  Check that the CPR valve is properly attached.

•  Check all air cells for leakage. This is easiest to do when the pump is set to 

“Care” 

  

(Auto-firm mode) and all cells are properly connected.

If the alarm still persists and the 

“        ”

 lamp does not switch off, the mattress should be 

replaced. Contact Järven Health Care for consultation and advice – information can be found 
on the back cover of this manual.

Содержание OptiCell Soft PLUS

Страница 1: ...1 USER GUIDE OptiCell Soft PLUS Constant low air pressure overlay mattress...

Страница 2: ...important to the user Risk of damage to mattress Important information denoted by this symbol must be observed in order not to damage the mattress or other material CE MARKING J rven Health Care hygi...

Страница 3: ...Categories 1 3 low risk to very high risk for patients within the weight range 30 230 kg The product is intended for use with a sheet between the patient s skin and the mattress surface The mattress...

Страница 4: ...01 EN 597 1 and EN 597 2 SYSTEM FEATURES The OptiCell Soft PLUS air mattress is constructed so that individual cells can be disconnected even while a patient is lying on the mattress The mattress has...

Страница 5: ...re We recommend that the mattress be kept rolled up in one of our optional transport bags The mattress should not be treated carelessly when stored as it risks being damaged Avoid exposing the mattres...

Страница 6: ...ny problems with the mattress s air pressure The test ensures that the patient does not bottom out How the hand check is performed The hand check should always be performed when the patient is positio...

Страница 7: ...on but the pump does not work contact J rven Healthcare The mattress does not inflate properly pressure too low Set the pressure to maximum check if the air hoses are pinched anywhere or if there is a...

Страница 8: ...mately 15 minutes When the lamp is lit the mattress is ready to use UNLOCK The panel can be locked to prevent unauthorised people from changing the system settings To unlock the panel press and hold t...

Страница 9: ...slation EU Directive 85 374 EEC Swedish Law SFS 1992 18 establishes the manufacturer s importer s or seller s statutory liability for any damage caused by the product The regulations apply to physical...

Страница 10: ...ent regarding pressure relief and must be conducted by staff with adequate knowledge and education Based on the patient s assessed status the appropriate mattress is chosen for the correct purpose and...

Страница 11: ...0 824 10 Email healthcare jarven se OptiCell Soft PLUS user guide All trademarks are the property of their respective owners J rven AB reserves the right to update this guide if necessary and or on fi...

Отзывы: