11

HAND CHECK TEST

Test the mattress’s function with a simple hand check.

To ensure that the mattress is providing the right therapeutic function, a simple hand check is 
performed. The test should be performed each time a mattress is taken into operation in order to 
detect any problems with the mattress’s air pressure. The test ensures that the patient does not 
“bottom out”.

How the hand check is performed.

The hand check should always be performed when the patient is positioned on the mattress. It 
is important to ask the patient not to try and lift their body during the test and to stay lying as 
usual. It is not uncommon for patients to lift their body in the belief that it is easier for the staff, 
but it may give an incorrect test result.

•  Fold away the sheets and unzip the mattress cover to access the cells.

•  Put one hand between two cells with the patient lying on the mattress. 

(See illustration)

.

Depending on the size of the hand, it may sometimes be sufficient to use 2-3 finger height.

•  Ensure that when the hand is in position, it fits between the bed base and the patient.

The important conclusion of the check is that there is space between the patient and the 
bed base.

If the hand check shows that the patient is “bottoming out”.

If the patient is “bottoming out” and no alarm has triggered on the pump, the comfort level

 

should be raised 

(position B on page 14)

. If it is already set to maximum and the patient does 

not weigh more than the recommended maximum weight, then the mattress and pump need 
troubleshooting by authorised personnel (see page 12).

Содержание OptiCell 2 NX

Страница 1: ...1 USER GUIDE OptiCell 2 NX Alternating pressure overlay air mattress ...

Страница 2: ...roduct 8 Receiving and unpacking 8 Storage 8 Automatic adjustment and alternating pressure 9 Static mode 9 Low pressure alarm 9 Air hose connection 10 Hand Check test 11 Inspection and troubleshooting 12 Cleaning 13 Pump overview OC 20 14 Specifications 15 Recycling 16 Other information 17 Product liability 17 Returns and complaints 17 Damage upon delivery 17 Pressure ulcer prevention work 18 Your...

Страница 3: ...G The CE marking signifies that OptiCell AB has the manufacturer s responsibility to ensure that this product meets the essential requirements of relevant European legislation This includes the provision of technical documentation and assurance which is maintained for at least five years after the product is placed on the market QUALITY Our quality standards and customer focus goals are realised b...

Страница 4: ...sment of the intended usage environment and then use the right product Please contact us if you have any questions In order to avoid damage the mattress should not come into contact with heat sources or fire such as burning cigarettes heat lamps or the like Make sure that you use the correct mattress size for the bed and the appropriate model for the patient Connect the product to the nearest situ...

Страница 5: ...use maintenance and improved safety It is very important to read this guide carefully before using the product Care personnel must be given access to this user guide and also be informed of the risks that may arise when using electrical equipment Training on the product can be provided on request Before using the product for the first time or when it is retrieved from a storage location Check the ...

Страница 6: ...ith the bi elastic surface layer reduces the risk of shear and other complications for the patient Fluid protection flap and waterproof zip on four sides with welded seams Anti slip underside which grips to bed base in addition to four fastening straps Four transportation handles to enable easy moving APPLICATIONS The OptiCell 2 NX is an alternating pressure overlay air mattress which is placed on...

Страница 7: ...he pump by three air hoses The red air hose positioned in the middle is used to fill the head cells situated under the patient s head It also fills the lower chambers of the standard cells and the ventilation cells This ensures that there is always air in the mattress no matter how the pump is set or if the pump should stop working for example due to power failure The outer two air hoses provide e...

Страница 8: ...in mind that a tall patient can weigh the same as a short patient but may need a different pressure This must always be carefully observed for each patient The biggest risk for the patient is to bottom out on the mattress The basic rule is that the part of the body to be unburdened must not be subject to pressure If the patient bottoms out the objective of the mattress is lost See page 11 for info...

Страница 9: ...t can be for instance during nursing care procedures or when the patient does not want to be disturbed by an alternating pressure Press the Static Dynamic button position E on page 14 to change to static air pressure from alternating air pressure The correct therapeutic effect is achieved with alternating air pressure in Dynamic mode Static air pressure constant low pressure should only be used wh...

Страница 10: ...e connector if the air hose is disconnected from the pump To minimise the risk of air leakage during hose connection check That the two hooks are not damaged and their clips are intact That the black o rings on the connector s three air hoses are not cracked or missing That no debris has entered the connector causing a bad connection or blocking the air hoses The air hose connector is fitted by a ...

Страница 11: ...it is easier for the staff but it may give an incorrect test result Fold away the sheets and unzip the mattress cover to access the cells Put one hand between two cells with the patient lying on the mattress See illustration Depending on the size of the hand it may sometimes be sufficient to use 2 3 finger height Ensure that when the hand is in position it fits between the bed base and the patient...

Страница 12: ...ir hoses for signs of damage or abnormal wear Check the air hose connector see page 10 INSPECTION OFTHE OPTITEX COVER Regularly check the cover for visible signs of damage or wear such as tears cracks holes stains or discolouration Check all sides including the underside TROUBLESHOOTING The pump does not work or does not start Make sure the electrical plug is properly inserted then turn on the pum...

Страница 13: ...ed at a minimum of 70 C for at least 10 minutes with unscented detergent according to standard BS EN 14065 2016 Textiles Laundry Processed Textiles Biocontamination Control System The OptiTex cover can be machine washed at a maximum of 95 C CLEANING THE CELLS AND AIR HOSES The cells can be washed separately or in groups using a damp cloth and mild detergent The air hoses can also be wiped with a w...

Страница 14: ...re are six comfort modes to choose from The patient s experience of lying on the mattress is affected by the comfort mode chosen Press this button to cycle through modes ALARM If the alarm sounds press this button to mute and reset The adjacent Low lamp will light up if the air pressure in the mattress is too low Check if necessary for any air leakage see page 12 The adjacent lamp marked will ligh...

Страница 15: ... static mode When the adjacent Dynamic lamp is lit the mattress is in alternating pressure mode and the cycle time is 10 minutes SPECIFICATIONS Pump OC 20 Specifications Size 28 0 x 11 0 x 20 5cm Weight 2 6 kg Pressure range 20 70mm Hg Cycle time 10 minutes System Specifications Classification Class II IPXO No AP APG Type BF Applied part Mattress Power rating 230V AC 50 Hz 1A 12W Fuse rating 250V ...

Страница 16: ...sposed of at a recycling station and sorted as Combustible waste For more information contact Järven HealthCare Järven Health Care takes environmental responsibility seriously We strive to always use materials and products with the least possible impact on the environment We are constantly working to develop and improve production techniques Järven Health Care is affiliated with the Swedish FTI an...

Страница 17: ... COMPLAINTS Any complaints regarding the product must be notified in writing to Järven Health Care The return of products must be done in accordance with our instructions Returning a product before a complaint case is processed will result in the case being considered closed and customer compensation will not be paid DAMAGE UPON DELIVERY Any damage of the product must be noted on the accompanying ...

Страница 18: ...sessment is made to identify the specific needs of the patient regarding pressure relief and must be conducted by staff with adequate knowledge and education Based on the patient s assessed status the appropriate mattress is chosen for the correct purpose and pressure relief needs Routinely change the patient s location and position on the mattress throughout the care period In addition regularly ...

Страница 19: ...19 YOUR NOTES If you require any further information or assistance please get in touch with us We will be happy to help you ...

Страница 20: ...n se OPTICELL 2 NX user guide All trademarks are the property of their respective owners Järven AB reserves the right to update this guide if necessary and or on finding any spelling or factual errors Additional copies can be ordered from Järven Healthcare Copying for internal use is allowed OptiCell 2 NX rev 19 12 00 ...

Отзывы: