19
̶
GO-2400M-PMCL / GO-2400C-PMCL
■
Exposure times
Exposure times vary depending on the Tap Geometry and CL Pixel Clock settings. Refer to the
following table.
Tap Geometry
CL Pixel Clock (MHz)
Max. exposure time value A
Longest
exposure time
[msec]
Shortest
exposure time
[usec]
1X2-1Y
37.125
Frame Period -342 [usec]
7999.658
79
1X2-1Y
74.25
Frame Period -171 [usec]
7999.829
40
1X2-1Y
84.85
Frame Period -162 [usec]
7999.838
38
1X3-1Y
37.125
Frame Period -252 [usec]
7999.748
59
1X3-1Y
74.25
Frame Period -114 [usec]
7999.886
27
1X3-1Y
84.85
Frame Period -100 [usec]
7999.900
24
1X4-1Y
37.125
Frame Period -172 [usec]
7999.828
40
1X4-1Y
74.25
Frame Period -86 [usec]
7999.914
20
1X4-1Y
84.85
Frame Period -81 [usec]
7999.919
19
1X8-1Y
37.125
Frame Period -88 [usec]
7999.912
21
1X8-1Y
74.25
Frame Period -63 [usec]
7999.937
15
1X8-1Y
74.25
Frame Period -81 [usec]
7999.919
19
1X8-1Y
84.85
Frame Period -63 [usec]
7999.937
15
1X8-1Y
84.85
Frame Period -81 [usec]
7999.919
19
Control via External Triggers
When Controlling the Exposure Time Using Specified Exposure Times
Configure the settings as follows.
Item
Setting value / selectable range
Trigger Selector (trigger operation)
Frame Start
Trigger Mode
On
Trigger Source (trigger signal source)
Any
Trigger Activation (trigger polarity)
Rising Edge (rising edge of input signal), Falling Edge (falling
edge of input signal)
Exposure Mode
Timed (control via exposure time)
Exposure Time
Varies depending on the Tap Geometry and CL Pixel Clock
settings. *
1
Exposure Auto
Off, Continuous
* 1 Max. value = [Acquisition Frame Rate] - Max. exposure time value A (“Exposure times” (page 19))
1
Set [Exposure Mode] to [Timed].
([Timed] is the default setting.)
2
Specify the exposure time in [Exposure Time].
The setting value for the exposure time can only be changed when [Exposure Auto] is set to [Off].
If [Exposure Auto] is set to [Continuous], temporarily set it to [Off] before changing the [Exposure
Time].
3
Set [Trigger Selector] to [Frame Start].
([Frame Start] is the default setting.)
4
Set [Trigger Mode] to [On].