ADwin-light-16
, manual version 2.2, December 2004
9
Inputs and Outputs
ADwin
5 Inputs and Outputs
Connections
The
ADwin-light-16
system is equipped with a connecting socket for USB or
Ethernet as well as with a 37-pin D-SUB socket. You will find on the D-SUB
socket (pin assignment next page) as follows:
– 8 analog inputs
– 2 analog outputs
– 6 digital inputs and outputs each
– 1 digital trigger input
– 2 impulse/event counters with 32 bit
– Output for power 5V;
L16-PCI
and
L16-cPCI
also ±12V
The design
L16-EXT
has an additional GND socket (see earth protection,
page 7), a power input socket and a manual on/off switch.
All inputs and outputs may only be operated according to the specifications
given (see Annex A-1: Technical Data). In case of doubt, ask the manufacturer
of the device to which you want to connect the
ADwin-light-16
system.
Open inputs can cause errors – above all in an environment which is not free
of any interferences. For your own safety, set unused inputs as close as pos-
sible to the D-SUB socket on a defined level (e.g. GND). Separate these inputs
also from open-ended lines.
Exception to this is the event input, which already has an internal pull-up resis-
tor (10 k
Ω
).
Front
Back
L16-PCI
L 1 6 - E U R O
and
L16-cPCI
L16-EXT
Fig. 5 – Connections at the
ADwin-light-16
system
ADwin
-I/O-CONNECTOR
ADwin
-L16
USB
USB
ADwin
I/O CONNECTOR
10-18VDC
ON
POWER
GND