JCL Service Manual
- www.cantudong,com
4-8 Function setting
Turn on the power while pressing key
G/N/SET
; the scale displays CHECK.
Use
MR
or
TARE
to step through all the available setting items.
Use
G/N/SET
to enter the parameter setting.
Then use
MR
or
TARE
to set the content values of setting items.
Press
M+/SAVE
to save the setting and return, or press
ZERO/ESC
to return without
saving.
Press
ZERO/ESC
to return normal weighing mode.
Note: Press and hold key
SMPL
and
G/N/SET
while powering on the scale. Press the passcode “11” and
press
M+/SAVE
to enter the advanced parameter setting.
5. Trouble shootings
5-1 Preliminary checks
Is the bubble of the level indicator at the center?
Is the battery electricity capacity ok?
Is power cord connected both with the scale and with the outlet properly?
Is the weighing pan the right one?
Is there anything under the weighing pan?
Is the scale at a flat and steady base?
Are there any vibrating, rotating, and/or reciprocating equipment around?
5-2 Problems and solutions