Chapter 5
– Software
32
Int32 PR6xxComm.CommBase.prDeviceGetTakt (ref int Takt1, ref int Takt2, ref int
Takt3, ref int Takt4)
Get previously configured TAKT table values. (These are saved in the SDK).
Parameters:
Takt1
Last exposure time in microseconds for bright light/color levels
Takt2
Last exposure time in microseconds for medium light/color levels
Takt3
Last exposure time in microseconds for low light/color levels
Takt4
Last exposure time in microseconds for very low light/color levels
Returns:
Int32 PR6xxComm.CommBase.prDeviceInstCalDue (ref string szDueDate)
Report the date and time when the instrument will be out of calibration.
Parameters:
szDueDate
Instrument calibration due date/time string
Returns:
A Int32 instrument operation result code: 0 = success, negative number is error code
Int32 PR6xxComm.CommBase.prDeviceLastMeasurementInfo (ref Int32 pixel, ref Int32
lightDark, ref double exposure, ref double brightness, ref Int32 cct, ref double
temperature)
Get additional information for last luminance or spectral measurement.
Parameters:
pixel
Measured pixel with highest A/D count.
lightDark
Average light minus dark A/D count for last measurement.
exposure
Exposure time of last measurement.
brightness
Luminance value returned from last measurement.
cct
Color correction temperature returned for las measurement.
temperature
CCD temperature for last measurement.
Returns:
A Int32 instrument operation result code: 0 = success, negative number is error code.
Содержание PHOTO RESEARCH A-TAKT V-7HS
Страница 2: ......
Страница 11: ...Chapter 2 System Overview 7...
Страница 18: ...Chapter 3 Theory of Operation 14...
Страница 47: ...43 FIGURE 13 A TKT V 7HS ENVELOPE DRAWING Front Mounting Hole...