![J-KEM Scientific Gemini-230 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/j-kem-scientific/gemini-230/gemini-230_manual_3079624038.webp)
38
Classes of J-KEM Scientific Temperature Controllers
J-KEM’s 200-Series, Apollo, and Gemini controllers are compatible with any size heating mantle from 5 ml
to 50 liter, and any 120 Vac oil bath (do not use with oil baths rated less than 120 Vac).
J-KEM’s 150-Series economy controllers are compatible with 500 ml heating mantles and larger, and 120
Vac oil baths (do not use with oil baths rated less than 120 Vac).
J-KEM’s 230 Vac controllers are compatible with any 230 Vac heating mantle 100ml and larger. They
are not compatible with any size oil bath.
Shown is a typical set up for a solution phase reaction.
1.
Place the reaction flask in the appropriate size heating mantle., then plug the power cord from the
mantle directly into the power outlet of the temperature controller.
2.
Place a stirring bar inside of the flask, in the solution. Place the flask on a magnetic stirrer and stir
the reaction for good heat transfer.
3.
Place a thermocouple in solution and make sure that at least the first ¼” (1/2” is better) is covered by
the fluid in the reaction flask.
Heater power cord
Thermocouple
in solution
100 ml Heating
mantle