Section 4:
Application Notes
Supplemental application notes on the following topics are available by contacting J-KEM.
Application
Note
Subject
AN1
How to heat oil baths with your controller. (Included in Appendix)
AN2
Changing the controllers thermocouple type.
AN3
Changing the heating outlet into a cooling outlet.
AN4
Using the controller for unattended fractional distillations.
AN5
Using the controller with inductive (motors, valves) loads.
4.1
Theory of How the Controller Works – Simply.
For the purpose of explaining how the controller
works, the example of a solution heated with a heating mantle is used. The principles are the same for all
heater types.
Power
ON
J-KEM
Scientific
Instruments for Science
Model 210
120 Vac, 840 watts
120 vac
Power to heater
Thermocouple
Heating mantle
Temperature of process
from the thermocouple
Output
Power Level
1-10 ml
10-100 ml
50-500 ml
300 ml-2 L
> 2 L
Heat
Off
8 0. 0
o
C
*
Figure 2
The controller, the heating mantle and the thermocouple form a closed loop feedback system. If the
process temperature is below the set point, the controller turns the heating mantle on and then monitors the
temperature rise of the solution. If a small rise results (indicating a large volume is being heated) the
controller sets internal parameters appropriate for heating large volumes. If a large rise in temperature
results, the controller responds by loading a set of parameters appropriate for heating small volumes. For
the controller to work ideally, information needs to travel instantaneously around the feedback loop. That
means that any power the controller applies to the heating mantle must reflect itself in an instantaneous
temperature rise of the solution and the thermocouple. Unfortunately, this type of instantaneous heat transfer
from the heating mantle to the solution to the thermocouple just doesn't occur. The delay time between when
power is applied to the heating mantle and when the solution rises in temperature; and also the converse,
when power is removed from the heating mantle and the solution temperature stops rising is the source of
most controller errors. The reason for this can be seen in a simple example.