background image

numbered from left to right:

• 

Sockets 1 (Common)

2 (NC)

, and 

3 (NO)

 are the connections for 

Limit 

Relay A

.

• 

Sockets 4 (Common)

5 (NC)

, and 

6 (NO)

 are for 

Limit Relay B

.

• 

Sockets 7 (Common)

8 (NC)

, and 

9 (NO)

 are for 

Warning Relay

.

• 

Sockets 10 (Common)

11 (NC)

, and 

12 (NO)

 are for 

System OK

.

6) An analog output for the carbon dioxide sensor and an additional analog output 

for the oxygen sensor are available from the terminal block. These outputs are 
non-isolated current sources that are referenced to the current output ground. 
The range of each current output is 4 to 20 mA and a maximum output voltage 
of 10 Volts. The oxygen and carbon dioxide concentrations that are equal to the 
outputs at 4 mA and 20 mA are normally set to the low and high limits of the 
sensitivity ranges of the two gases, respectively: For example, the 4 mA current 
output of the oxygen sensor is set to 0%, and the 20 mA current output is set to 
100%.

• 

Socket 13 is the 

Ground

 for the both current outputs.

• 

Socket 14 is the 

oxygen current output

.

• 

Socket 15 is the 

carbon dioxide current output

.

GA-200B: Rear Panel

GA-200B Rear Panel

The rear panel of the GA-200B has the following features: a switch, a power input, 
an RS-232 serial port for direct connection to a computer, an oxygen sensor 
output, a carbon dioxide sensor output, and a gas outlet.

1) Power for the GA-200B is supplied by an external power supply. The power 

supply is connected to the unit through the connector labeled 

12V

1.5A

. Press 

the side of the power switch labeled 

I

 to turn on the GA-200B. Press 

O

 on the 

switch when it is time to turn the unit off.

2) The GA-200B can be interfaced to a computer through a serial (

RS-232

) data 

stream. The default data format is 

9600 baud

8 bit

, and 

no parity

 with 

alternate baud rates of 2400, 4800, 19200, and 38400. The signals present are 

RXD

TXD

, and ground through the 9-pin D connector on the back panel of the 

unit.

GA-200 Gas Analyzer 

iWorx Systems, Inc.

www.iworx.com

Содержание GA-200 Series

Страница 1: ...oxide in a gas a sample is pumped from the input port on the front panel of the GA 200 through the sample cell and out an exhaust port on the back of the unit Gas samples are drawn into the unit throu...

Страница 2: ...to a computer through a serial RS 232 data stream The default data format is 9600 baud 8 bit and no parity with alternate baud rates of 2400 4800 19200 and 38400 The signals present are RXD TXD and g...

Страница 3: ...the 20 mA current output is set to 100 Socket 13 is the Ground for the both current outputs Socket 14 is the oxygen current output Socket 15 is the carbon dioxide current output GA 200B Rear Panel GA...

Страница 4: ...the surface of a detector The output of the sensor is inversely proportional to the concentration of oxygen in the sample because the amount of light reaching the detector decreases as the concentrati...

Страница 5: ...en the output of the carbon dioxide sensor which is located on the terminal block on the rear panel of the analyzer and the BNC input of Channel 1 on the IWX 214 Red wire into socket 15 and black wire...

Страница 6: ...lve on the regulator of the cylinder providing the second gas sample Adjust the flow rate to low Make sure the flow rate from the gas cylinder is greater than the flow rate on the gas analyzer 2 While...

Страница 7: ...cribe toolbar so that two cursors appear on the Main window Place one cursor on the section of data recorded when gas analyzer was collecting a sample of room air and the second cursor on the section...

Страница 8: ...the GA 200 gas analyzer 5 Notice that the voltages from the positions between the cursors are automatically entered into the value equations Enter the two concentrations of carbon dioxide measured fr...

Страница 9: ...be disinfected An external barrier filter on the inlet port must be used to prevent contamination of the sample circuit 4 If the GA 200 is used to collect samples of inspired and expired air during b...

Страница 10: ...zer 12VDC 1 5A Requirements of the External Power Supply 95 250VAC 47 63 Hz Dimensions 190 x 76 x 280mm WxHxD Weight Analyzer 2 3kg Power Supply 0 7kg Storing Before storing the GA 200 connect an exte...

Отзывы: