iWorX GA-200 Series Скачать руководство пользователя страница 4

3) The signal from the oxygen sensor can be recorded from the BNC output 

labeled 

OXYGEN SENSOR

0-1V=0-100%

. The amplitude of the signal ranges 

from 0 to 1 Volt DC which corresponds to a range of 0 to 100% oxygen, or 1mV 
for each 0.01% oxygen. The ground of this output is common to the instrument 
case and the power supply.

4) The signal from the carbon dioxide sensor can be recorded from the BNC 

output labeled 

CO2 SENSOR, 0.8-4V=0- 10%

. The amplitude of the signal 

ranges from 0.8 to 4 Volt DC which corresponds to a range of 0 to 10% carbon 
dioxide, or 1mV for each 0.01% oxygen.

Items Provided With the GA-200 Gas Analyzer

External filter with Luer–Lock fittings
Nafion sampling tube with Luer-Lock fittings
Power supply
GA-200A: One BNC-BNC cable, one BNC- Two wire cable
GA-200B: Two BNC-BNC cables

How the GA-200 Works

The oxygen concentration of the gas in the sample cell is measured using laser 
diode absorption technology. The laser diode in the oxygen sensor produces light 
at a wavelength (760 nanometers) that is absorbed by oxygen. The light passes 
through the gas pumped into the sample cell and onto the surface of a detector. 
The output of the sensor is inversely proportional to the concentration of oxygen in 
the sample because the amount of light reaching the detector decreases as the 
concentration of oxygen in the sample increases. The GA-200 has a very fast 
response time that enables breath to breath analysis of gas concentrations 
because the unit analyzes the gas sample every 10 milliseconds, or 100 times per 
second. During each measurement interval, the analyzer is zeroed automatically 
by electronic tuning of the laser to a wavelength not absorbed by oxygen.

How to Use the GA-200 Gas Analyzer

Prior to use, users must perform a two-point calibration of the oxygen and carbon 
dioxide sensors.

Note: Warm up the GA-200 for at least 30 minutes. The input air filter must be 
connected to avoid damaging the sensors.

Calibrating the O

2

 and CO

2

 Channels

The outputs of the oxygen and carbon dioxide sensors of the GA-200 are voltages 
that are proportional to the concentrations of the gases being measured by the 
analyzer. To determine the volumes of oxygen consumed and carbon dioxide 
produced during metabolic testing, the voltage outputs of the sensors need to be 
converted, by the recording software, to the percentages of these gases in the 

GA-200 Gas Analyzer 

iWorx Systems, Inc.

www.iworx.com

Содержание GA-200 Series

Страница 1: ...oxide in a gas a sample is pumped from the input port on the front panel of the GA 200 through the sample cell and out an exhaust port on the back of the unit Gas samples are drawn into the unit throu...

Страница 2: ...to a computer through a serial RS 232 data stream The default data format is 9600 baud 8 bit and no parity with alternate baud rates of 2400 4800 19200 and 38400 The signals present are RXD TXD and g...

Страница 3: ...the 20 mA current output is set to 100 Socket 13 is the Ground for the both current outputs Socket 14 is the oxygen current output Socket 15 is the carbon dioxide current output GA 200B Rear Panel GA...

Страница 4: ...the surface of a detector The output of the sensor is inversely proportional to the concentration of oxygen in the sample because the amount of light reaching the detector decreases as the concentrati...

Страница 5: ...en the output of the carbon dioxide sensor which is located on the terminal block on the rear panel of the analyzer and the BNC input of Channel 1 on the IWX 214 Red wire into socket 15 and black wire...

Страница 6: ...lve on the regulator of the cylinder providing the second gas sample Adjust the flow rate to low Make sure the flow rate from the gas cylinder is greater than the flow rate on the gas analyzer 2 While...

Страница 7: ...cribe toolbar so that two cursors appear on the Main window Place one cursor on the section of data recorded when gas analyzer was collecting a sample of room air and the second cursor on the section...

Страница 8: ...the GA 200 gas analyzer 5 Notice that the voltages from the positions between the cursors are automatically entered into the value equations Enter the two concentrations of carbon dioxide measured fr...

Страница 9: ...be disinfected An external barrier filter on the inlet port must be used to prevent contamination of the sample circuit 4 If the GA 200 is used to collect samples of inspired and expired air during b...

Страница 10: ...zer 12VDC 1 5A Requirements of the External Power Supply 95 250VAC 47 63 Hz Dimensions 190 x 76 x 280mm WxHxD Weight Analyzer 2 3kg Power Supply 0 7kg Storing Before storing the GA 200 connect an exte...

Отзывы: